Assessment of collimators in radium-223 imaging with channelized Hotelling observer: a simulation study
Akihiko Takahashi1, Shingo Baba2, Masayuki Sasaki3
1Division of Medical Quantum Science, Department of Health Sciences, Kyushu University, Fukuoka, Japan. takahsr@hs.med.kyushu-u.ac.jp.
Objective:
Radium-223 (223Ra) is used in unsealed radionuclide therapy for metastatic bone tumors. The aim of this study is to apply a computational model observer to 223Ra planar images, and to assess the performance of collimators in 223Ra imaging.
Methods:
The 223Ra planar images were created via an in-house Monte Carlo simulation code using HEXAGON and NAI modules. The phantom was a National Electrical Manufacturers Association body phantom with a hot sphere. The concentration of the background was 55 Bq/mL, and the sphere was approximately 1.5-20 times that of the background concentration. The acquisition time was 10 min. The photopeaks (and the energy window) were 84 (full width of energy window: 20%), 154 (15%), and 270 keV (10%). Each 40 images, with and without hot concentration, were applied to a three-channel difference-of-Gaussian channelized Hotelling observer (CHO), and the signal-to-noise ratio (SNR) of the hot region was calculated. The images were examined using five different collimators: two low-energy general-purpose (LEGP), two medium-energy general-purpose (MEGP), and one high-energy general-purpose (HEGP) collimators.
Results:
The SNR value was linearly proportional to the contrast of the hot region for all collimators and energy windows. The images of the 84-keV energy window with the MEGP collimator that have thicker septa and larger holes produced the highest SNR value. The SNR values of two LEGP collimators were approximately half of the MEGP collimators. The HEGP collimator was halfway between the MEGP and LEGP. Similar characteristics were observed for other energy windows (154, 270 keV). The SNR value of images captured via the 270-keV energy window was larger than 154-keV, although the sensitivity of the 270-keV energy window is lower than 154-keV. The results suggested a positive correlation between the SNR value and the fraction of unscattered photons.
Conclusions:
The SNR value of CHO reflected the performance of collimators and was available to assess and quantitatively evaluate the collimator performance in 223Ra imaging. The SNR value depends on the magnitudes of unscattered photon count and the fraction of unscattered photon count. Consequently, in this study, MEGP collimators performed better than LEGP and HEGP collimators for 223Ra imaging.
Related Concept Videos
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...
Naturalistic Observations
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...
Observational Learning
Actor-Observer Effect
Imaging Studies VII: Vascular Imaging