Magnetic Intramedullary Lengthening Nails and MRI Compatibility
Charles Gomez1, Scott Nelson1, Joshua Speirs2
1Departments of Orthopedics.
Background:
Magnetic intramedullary nails (IMNs) are fully implantable lengthening devices that became available in the United States in 2011 for the correction of limb length discrepancies. This device represents a major advancement in the field of limb lengthening surgery as it is typically tolerated better than external fixation. Unlike traditional IMNs, surgeons recommend routine removal following limb lengthening. One such reason involves patient safety as it pertains to magnetic resonance imaging (MRI). Theoretical concerns with MRI exposure include implant migration, implant heating, and involuntary elongation of the lengthening mechanism. Our study seeks to investigate the effects of MRI on intramedullary magnetic lengthening nails.
Methods:
Twenty-five intramedullary magnetic nails were studied. One nail was placed within the magnetic field to measure maximum magnetic force. Nails were then scanned using standard knee MRI protocols, 12 in 3 T and 12 in 1.5 T MRI scanners. The following parameters were measured: (1) distraction of the implants after MRI exposure, (2) temperature before and after MRI, and (3) internal distraction force before and after MRI.
Results:
Maximum magnetic force was found to be 2 lbs. There was no involuntary distraction of the implants after MRI. Temperature increase of 3.3°C was found in the femoral nails and 3.6°C in the tibial nails that were exposed to 3 T MRI. This increase did not reach or exceed physiological temperature of 37°C. Distraction force was reduced by 61.7% in the femoral nails and 89.6% in the tibial nails after subjected to 3 T MRI. There was no reduction in distraction force after exposure to 1.5 T MRI.
Conclusions:
Recommendations for routine removal of magnetic IMNs for safety concerns should be reconsidered. Exposure to 3 T MRI should be avoided in patients who are still undergoing lengthening or with plans for future lengthening with magnetic IMNs.
Clinical Relevance:
To assess patient safety and implant function after magnetic IMNs have been exposed to MRI.
More Related Videos
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
07:34Functional MRI in Conjunction with a Novel MRI-compatible Hand-induced Robotic Device to Evaluate Rehabilitation of Individuals Recovering from Hand Grip Deficits
Published on: November 23, 2019
Related Concept Videos
Accessory Structures of the Skin: Nails
The main components of a nail include the following.
Nail Plate: The nail plate is the visible portion of the nail that extends beyond the fingertips or toes. It is a hard, translucent...
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetism
An individual magnetic pole cannot be isolated. No matter how small, every piece of a magnet contains a north pole and a south...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Magnetic Flux
Suppose a surface is divided into elements of area dA. For each element, the component of the magnetic field that is normal to the...