Related Experiment Video
Updated: Feb 7, 2026

Technical Aspects of the Mouse Aortocaval Fistula
Published on: July 11, 2013
Technical Aspects of Using the AngioVac System for Thrombus Aspiration From the Ascending Aorta
Nikolaos Tsilimparis1, Konstantinos Spanos1, E Sebastian Debus1
11 German Aortic Center Hamburg, Department of Vascular Medicine, University Heart Center, Hamburg, Germany.
Purpose:
To present the technique for removing mural thrombus from the ascending aorta using the AngioVac System.
Technique:
The technique is demonstrated in a 66-year-old woman who presented with free-floating mural thrombus in the ascending aorta and was considered unsuitable for either open (comorbidities) or endovascular approaches (high risk of stroke). Because of the free-floating thrombus, the Angiovac system was suggested, although it is approved for only the venous system. The technique was adapted for the arterial system with the 2 access points being (1) the proximal left subclavian artery with a 10-mm conduit for device access and (2) the right femoral vein. The AngioVac cannula and the reinfusion cannula were inserted into the artery and vein, respectively. The extracorporeal bypass circuit was created, and the carotid arteries were clamped during aspiration (<1 minute). The mural thrombus was aspirated successfully. The patient had an uneventful postoperative course with no signs of free thrombus on the postoperative or follow-up computed tomography angiograms.
Conclusion:
The use of the AngioVac System seems to be a feasible technique for aspiration of thrombus from the ascending aorta. Expanding this therapeutic option for patients unsuitable for open or endovascular repair may be proven efficient in the near future.
Related Concept Videos
The Aorta
The average diameter of the aorta is approximately 2-3 cm, but the size can vary depending on the section of the aorta and the individual's age, sex, and body size. The aorta is...
The Arch of Aorta
Encircling the heart, the coronary arteries form a ring-like structure before...
Thoracic Aorta
Abdominal Aorta
The celiac trunk, a singular artery, divides into the left gastric artery, which...
Quantitative Aspects of Drug-Receptor Interaction
Anatomy of the Intestines
Small Intestines
The small intestine is an ~7 meter-long tube with an inner diameter of just 2.5 cm. Since most nutrients are absorbed here, the inner lining of the...