Acceptability, Accuracy, and Safety of Disposable Transnasal Capsule Endoscopy for Barrett's Esophagus Screening
Sarmed S Sami1, Prasad G Iyer2, Prachi Pophali2
1Barrett's Esophagus Unit, Division of Gastroenterology & Hepatology, Mayo Clinic, Rochester, Minnesota; Division of Surgery and Interventional Science, University College London, London, United Kingdom.
Background & Aims:
Screening for Barrett's esophagus (BE) with conventional esophagogastroduodenoscopy (C-EGD) is expensive. We assessed the performance of a clinic-based, single use transnasal capsule endoscope (EG Scan II) for the detection of BE, compared to C-EGD as the reference standard.
Methods:
We performed a prospective multicenter cohort study of patients with and without BE recruited from 3 referral centers (1 in the United States and 2 in the United Kingdom). Of 200 consenting participants, 178 (89%) completed both procedures (11% failed EG Scan due to the inability to intubate the nasopharynx). The mean age of participants was 57.9 years and 67% were male. The prevalence of BE was 53%. All subjects underwent the 2 procedures on the same day, performed by blinded endoscopists. Patients completed preference and validated tolerability (10-point visual analogue scale [VAS]) questionnaires within 14 days of the procedures.
Results:
A higher proportion of patients preferred the EG Scan (54.2%) vs the C-EGD (16.7%) (P < .001) and the EG Scan had a higher VAS score (7.2) vs the C-EGD (6.4) (P = .0004). No serious adverse events occurred. The EG Scan identified any length BE with a sensitivity value of 0.90 (95% CI, 0.83-0.96) and a specificity value of 0.91 (95% CI, 0.82-0.96). The EG Scan identified long segment BE with a sensitivity value of 0.95 and short segment BE with a sensitivity values of 0.87.
Conclusions:
In a prospective study, we found the EG Scan to be safe and to detect BE with higher than 90% sensitivity and specificity. A higher proportion of patients preferred the EG Scan to C-EGD. This device might be used as a clinic-based tool to screen populations at risk for BE. ISRCTN registry identifier: 70595405; ClinicalTrials.gov no: NCT02066233.
Related Concept Videos
Barrett Esophagus-I: Introduction
This constant acid exposure transforms the esophagus's pink mucosal lining (stratified squamous epithelium) into a type of lining more similar...
Barrett Esophagus-II: Clinical Manifestations and Management
To diagnose Barrett's esophagus, healthcare providers often recommend an endoscopy for those showing symptoms of acid reflux. The procedure entails...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures V: ERCP
Patient...
Upper GI Series: Barium Swallow
Purpose and Procedure
Patients undergoing this procedure ingest a liquid containing barium sulfate with a chalky...