Impact of low-kVp scan technique on oral contrast density at abdominopelvic CT
Douglas H Sheafor1, Mark D Kovacs1, Philip Burchett1
1Department of Radiology, Medical University of South Carolina, 96 Jonathan Lucas Street, MSC 323, 210 CSB, Charleston, SC, 29425-3230, USA.
Purpose:
To assess quantitative and qualitative effects of kVp on oral contrast density.
Materials And Methods:
Three readers retrospectively reviewed 100 CT scans performed at a range of high- and low-energy settings, independently determining their preferred window and level settings for evaluation of the oral-contrast-opacified bowel. Contrast density was also assessed quantitatively in the stomach, jejunum, and ileum. Subsequently, a range of oral contrast dilutions were imaged at varying kVp's in a commercially available CIRS tissue equivalent phantom model.
Results:
In the retrospective patient study, mean oral contrast density increased significantly in the ileum compared to the jejunum (455.2 and 308.8 HU, respectively, p < 0.01). Similar findings were seen in patients regardless of patients' weight. Mean oral contrast density was higher on lower-energy scans, requiring more window/level adjustment. An oral contrast iodine concentration of 5.82-7.77 mg I/mL most closely approximated a target oral contrast density of 200 HU.
Conclusions:
Oral contrast density is strongly influenced by kVp, supporting use of more dilute oral contrast when using lower-kVp techniques.
Related Concept Videos
Leaky Scanning
Voltammetric Techniques: Linear-Scan (E vs Time)
Impact of Groups on Groups
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Oral Cavity
Teeth: The teeth are the hardest structures in our bodies. Humans have two sets of teeth throughout their lifetime: deciduous (baby) teeth and permanent teeth. Each tooth consists of several parts: the crown (visible part), the root (embedded in the jaw), enamel (hard outer...
Density