Beam hardening correction tool improves the diagnosis of incipient caries lesions in Micro-CT images
Karla Rovaris1, Saulo Leonardo Sousa Melo2, Polyane Mazucatto Queiroz1
1Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas, Piracicaba, Brazil.
Objective:
To evaluate the effect of Beam Hardening Correction tool (BHC) on Micro-CT images used in the diagnosis of incipient caries lesions.
Methods:
Six posterior teeth, with 6 sound proximal surfaces and 6 natural white spot lesions, were scanned using a SkyScan1174 device. Each tooth was imaged 16 times in various acquisition parameters, and reconstructed 3 times applying different levels of BHC (0, 15 and 30%). The 288 datasets were analyzed by three examiners. Gold-standard was obtained by microscopic analyses under polarized light. Area under the ROC curve was obtained for each protocol and the BHC parameter was isolated to individually observe the influence of such parameters of reconstruction.
Results:
Microtomographic images were accurate for the diagnosis of incipient caries lesions even when the BHC tool was not applied. Accuracy was perfect in 50 % of the protocols without BHC. When adding the artefact correction, the number of protocols that reached perfect accuracy increased to 81.25%. No difference was found between the protocols that used 15 or 30% BHC. Statistically significant differences were found only in four comparisons, and on those the images that had BHC were more accurate. Two non-BHC protocols reached accuracy scores of 0.653 and 0.881, and were improved to perfect accuracy with BHC application.
Conclusion:
The use of the BHC tool should be encouraged when using Micro-CT to detect incipient caries lesions, especially when low-quality parameters aiming less acquisition and processing time are used.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Distance Corrections
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Power Factor Correction