Third generation dual-energy CT with 80/150 Sn kV for head and neck tumor imaging
Saravanabavaan Suntharalingam1, Elena Stenzel1, Axel Wetter1
1Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
Background:
Dual-energy CT (DECT) provides additional image datasets which enable improved tumor delineation or reduction of beam hardening artifacts in patients with head and neck squamous cell carcinoma (SCC).
Purpose:
To assess radiation dose and image quality of third-generation DECT of the head and neck in comparison to single-energy CT (SECT).
Material And Methods:
Thirty patients with SCC who underwent both SECT (reference tube voltage 120 kVp) and DECT (80/150 Sn kVp) of the head and neck region for staging were retrospectively selected. Attenuation measurements of the sternomastoid muscle, internal jugular vein, submandibular gland and tongue were compared. Image noise was assessed at five anatomic levels. Subjective image quality was evaluated by two radiologists in consensus.
Results:
CTDIvol was 55% lower with DECT (4.2 vs. 9.3 mGy; P = 0.002). Median image noise was equal or lower in DECT at all levels (nasopharynx: 3.9 vs. 5.8, P < 0.0001; floor of mouth: 3.6 vs. 4.5, P = 0.0002; arytenoids: 3.6 vs. 3.1, P = 0.096; lower thyroid: 4.4 vs. 5.7, P = 0.002; arch of aorta: 5.6 vs. 6.5, P = 0.001). Attenuation was significantly lower in DECT ( P < 0.05). Subjective image analysis revealed that DECT is equal or superior to SECT with regard to overall image quality (nasopharynx: 5 vs. 5, P = 1; floor of mouth: 5 vs. 5, P = 0.0041; arytenoids: 5 vs. 5, P = 0.6; lower thyroid: 5 vs. 3, P < 0.0001; arch of aorta: 5 vs. 4, P < 0.0001).
Conclusion:
Head and neck imaging with third-generation DECT can reduce radiation dose by half compared to SECT, while maintaining excellent image quality.
More Related Videos
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
08:04Monitoring Tumor Metastases and Osteolytic Lesions with Bioluminescence and Micro CT Imaging
Published on: April 14, 2011
Related Concept Videos
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
What is Energy?