The validity of adenosine triphosphate measurement in detecting endoscope contamination
C E McCafferty1, D Abi-Hanna2, M J Aghajani1
1Ingham Institute for Applied Medical Research, Sydney, NSW, Australia; Western Sydney University, School of Medicine, Sydney, NSW, Australia.
Background:
Endoscopic procedures are vital to gastrointestinal disease diagnosis and management, but risk infection transmission. In Australia, endoscopes undergo monthly-to-quarterly microbiological testing, to prevent patient infection. Endoscopes are used more frequently, meaning contamination may not be detected by this surveillance before infection transmission occurs.
Aim:
To evaluate the use of adenosine triphosphate (ATP) measurement, alongside standard microbiological cultures, in detecting endoscope contamination before high-level disinfection. Using these results, we also aimed to confirm the efficacy of manual cleaning in reducing levels of ATP and cfu/mL.
Methods:
Seventeen in-clinical-use gastroscopes and 24 in-clinical-use colonoscopes from the Liverpool Hospital Endoscopy unit were sampled across three separate cleaning stages before high-level disinfection. Colony counts and ATP measurements were then performed on these samples.
Findings:
The correlation between the cfu/mL and RLU of samples collected from colonoscopes was 0.497 (95% confidence interval: 0.28-0.66; P < 0.0001). The correlation between cfu/mL and RLU for samples collected from gastroscopes was 0.377 (0.08-0.61; P = 0.0138). RLU and cfu/mL values were shown to fall significantly (P < 0.005) following precleaning and manual cleaning.
Conclusion:
There was a significant correlation between ATP and cfu/mL measured from samples collected before high-level disinfection. Precleaning and manual cleaning were shown to reduce ATP and microbiological load significantly. ATP measurement can be performed within minutes with little training and produces results that are easy to interpret. These findings warrant further research on the utility of ATP measurement as a screening tool for detecting endoscope contamination after high-level disinfection.
Insights
Adenosine triphosphate (ATP) measurement correlates with microbiological cultures for detecting endoscope contamination. This rapid ATP test can help ensure endoscope safety by verifying cleaning effectiveness before disinfection.
Area of Science:
- Gastroenterology
- Infection Control
- Medical Device Reprocessing
Background:
- Endoscopic procedures are crucial for diagnosing gastrointestinal diseases but carry a risk of infection transmission.
- Current microbiological testing of endoscopes in Australia occurs monthly to quarterly, potentially lagging behind frequent use and increasing contamination risks.
- The study addresses the need for more timely detection of endoscope contamination.
Purpose of the Study:
- To assess the utility of adenosine triphosphate (ATP) measurement alongside standard microbiological cultures in identifying endoscope contamination prior to high-level disinfection.
- To confirm the effectiveness of manual cleaning procedures in reducing both ATP levels and microbial load (cfu/mL).
Main Methods:
- Seventeen gastroscopes and 24 colonoscopes from a clinical endoscopy unit were sampled at three distinct cleaning stages before high-level disinfection.
- Both colony counts (cfu/mL) and ATP measurements (relative light units - RLU) were performed on the collected samples.
Main Results:
- A significant positive correlation was found between ATP (RLU) and microbiological load (cfu/mL) for both colonoscopes (r=0.497, P < 0.0001) and gastroscopes (r=0.377, P = 0.0138).
- Both ATP and cfu/mL values decreased significantly (P < 0.005) after precleaning and manual cleaning stages.
- These findings indicate that cleaning procedures effectively reduce bioburden.
Conclusions:
- ATP measurement demonstrates a significant correlation with microbiological cultures for detecting contamination on endoscopes before high-level disinfection.
- Precleaning and manual cleaning significantly reduce both ATP levels and overall microbiological load.
- ATP measurement offers a rapid, easily interpretable screening tool for endoscope cleanliness, warranting further investigation for routine use.
More Related Videos
07:50Detection and Removal of Nuclease Contamination During Purification of Recombinant Prototype Foamy Virus Integrase
Published on: December 8, 2017
09:07Author Spotlight: Advancing Endoscopic Ossiculoplasty – Techniques, Innovations, and Practical Guidance for Clinical Integration
Published on: January 26, 2024
Related Concept Videos
Reliability and Validity
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: