The validity of adenosine triphosphate measurement in detecting endoscope contamination

C E McCafferty1, D Abi-Hanna2, M J Aghajani1

  • 1Ingham Institute for Applied Medical Research, Sydney, NSW, Australia; Western Sydney University, School of Medicine, Sydney, NSW, Australia.

Abstract

Insights

Adenosine triphosphate (ATP) measurement correlates with microbiological cultures for detecting endoscope contamination. This rapid ATP test can help ensure endoscope safety by verifying cleaning effectiveness before disinfection.

Area of Science:

  • Gastroenterology
  • Infection Control
  • Medical Device Reprocessing

Background:

  • Endoscopic procedures are crucial for diagnosing gastrointestinal diseases but carry a risk of infection transmission.
  • Current microbiological testing of endoscopes in Australia occurs monthly to quarterly, potentially lagging behind frequent use and increasing contamination risks.
  • The study addresses the need for more timely detection of endoscope contamination.

Purpose of the Study:

  • To assess the utility of adenosine triphosphate (ATP) measurement alongside standard microbiological cultures in identifying endoscope contamination prior to high-level disinfection.
  • To confirm the effectiveness of manual cleaning procedures in reducing both ATP levels and microbial load (cfu/mL).

Main Methods:

  • Seventeen gastroscopes and 24 colonoscopes from a clinical endoscopy unit were sampled at three distinct cleaning stages before high-level disinfection.
  • Both colony counts (cfu/mL) and ATP measurements (relative light units - RLU) were performed on the collected samples.

Main Results:

  • A significant positive correlation was found between ATP (RLU) and microbiological load (cfu/mL) for both colonoscopes (r=0.497, P < 0.0001) and gastroscopes (r=0.377, P = 0.0138).
  • Both ATP and cfu/mL values decreased significantly (P < 0.005) after precleaning and manual cleaning stages.
  • These findings indicate that cleaning procedures effectively reduce bioburden.

Conclusions:

  • ATP measurement demonstrates a significant correlation with microbiological cultures for detecting contamination on endoscopes before high-level disinfection.
  • Precleaning and manual cleaning significantly reduce both ATP levels and overall microbiological load.
  • ATP measurement offers a rapid, easily interpretable screening tool for endoscope cleanliness, warranting further investigation for routine use.

Related Concept Videos

Reliability and Validity01:29

Reliability and Validity

Reliability and validity are two important considerations that must be made with any type of data collection. Reliability refers to the ability to consistently produce a given result. In the context of psychological research, this would mean that any instruments or tools used to collect data do so in consistent, reproducible ways.
14.1K
Contaminants and Errors01:16

Contaminants and Errors

Effective sample preparation is crucial for accurate and reliable laboratory analysis. During this process, two significant sources of error can arise: concentration bias from improper sample splitting and contamination caused by methods used to reduce particle size, such as grinding or homogenization. Identifying and minimizing these potential errors is crucial to ensuring the validity of the analysis.
Another key consideration is determining the appropriate number of samples required to...
378
Data Validation01:15

Data Validation

Method validation is a crucial process in analytical chemistry designed to confirm that a given method consistently produces reliable and high-quality results. This process is essential when a method is applied to different sample matrices or when procedural modifications are made, ensuring that the results meet acceptable standards across various applications.
Key parameters for method validation include:
2.0K
Data Validation01:03

Data Validation

Data validation is an essential part of a comprehensive assessment. Validation is confirming or verifying and opening the door to gathering more assessment data as it clarifies vague or unclear data. The process of checking and verifying the collected information is called data validation. The primary purpose of data validation is to ensure data is as free from error, bias, and misinterpretation as possible.
Nursing assessment guides are generally based on holistic models rather than medical...
6.8K
Endoscopic Studies II: Thoracocentesis01:26

Endoscopic Studies II: Thoracocentesis

Thoracentesis(Thoracocentesis), commonly known as pleural tap, is a medical procedure where a 22 gauge needle is inserted into the pleural space, the area between the lung and chest wall. This procedure is commonly performed to diagnose or treat various respiratory disorders.
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
1.5K
Endoscopic Procedures I: Esophagogastroduodenoscopy01:29

Endoscopic Procedures I: Esophagogastroduodenoscopy

An Esophagogastroduodenoscopy (EGD) is a diagnostic procedure in which an endoscopist uses a flexible, lighted endoscope to visualize the upper gastrointestinal (GI) tract. The procedure includes visualizing the oropharynx, esophagus, stomach, and the first part of the small intestine, the duodenum.
During an EGD, the endoscope can be used to:
1.2K