New Volumetric Analysis Method for Stent Expansion and its Correlation With Final Fractional Flow Reserve and
Daisuke Nakamura1, William Wijns2, Matthew J Price3
1Harrington Heart and Vascular Institute, University Hospitals Cleveland Medical Center, Case Western Reserve University, Cleveland, Ohio.
Objectives:
This study sought to compare conventional methodology (CM) with a newly described optical coherence tomography (OCT)-derived volumetric stent expansion analysis in terms of fractional flow reserve (FFR)-derived physiology and device-oriented composite endpoints (DoCE).
Background:
The analysis of coronary stent expansion with intracoronary imaging has used CM that relies on the analysis of selected single cross-sections for several decades. The introduction of OCT with its ability to perform semiautomated volumetric analysis opens opportunities to redefine optimal stent expansion.
Methods:
A total of 291 lesions treated with post-stent OCT and FFR were enrolled. The expansion index was calculated by using a novel volumetric algorithm and was defined as: ([actual lumen area / ideal lumen area] × 100) for each frame of the stented segment. The minimum expansion index (MEI) was defined as the minimum value of expansion index along the entire stented segment. MEI and conventional lumen expansion metrics were compared for the ability to predict post-stent low FFR (<0.90) and DoCE at 1 year.
Results:
There was a stronger correlation between MEI and final FFR, compared with CM and final FFR (r = 0.690; p < 0.001) versus (r = 0.165; p = 0.044). MEI was significantly lower in patients with DoCE than those without DoCE (72.18 ± 8.23% vs. 81.48 ± 11.03%; p < 0.001), although stent expansion by CM was similar between patients with and without DoCE (85.05 ± 22.19% and 83.73 ± 17.52%; p = 0.858), respectively.
Conclusions:
OCT analysis of stent expansion with a newly described volumetric method, but not with CM, yielded data that were predictive of both an acute improvement in FFR-derived physiology and DoCE.
Related Concept Videos
Inverse z-Transform by Partial Fraction Expansion
To begin the process, the poles of the function are identified and the function is...
Correlations
Correlation and Causation
Correlation versus Causation
If the dependent variable increases or decreases when the independent variable increases, there is a positive or negative...
Correlation
Two variables, for example, a and b, are said to be positively correlated if both variables move in the same direction. In other words, a positive correlation exists between two variables, a and b, if:
Predicting Reaction Outcomes
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...