Related Experiment Video
Updated: Feb 6, 2026

A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
Complication and Failures of Central Vascular Access Device in Adult Critical Care Settings
Mari Takashima1, Jessica Schults2, Gabor Mihala1,3,4
1Alliance for Vascular Access Teaching and Research (AVATAR) group, Menzies Health Institute Queensland, Griffith University, Nathan, QLD, Australia.
Objectives:
To examine the proportion and rate of central venous access device failure and complications across central venous access device types in adult intensive care.
Data Sources:
A systematic search was undertaken in the electronic databases Cochrane Central Register of Controlled Trials, Embase, U.S. National Library of Medicine National Institutes of Health, and Cumulative Index to Nursing and Allied Health in September 2017.
Study Selection:
Included studies were of observational (prospective and retrospective) or interventional design and reported central venous access device failure and complications in adult ICU settings. Studies were excluded if they were published prior to November 2006 or not reported in English. Two reviewers independently screened articles, assessed eligibility, extracted data, and assessed risk of bias.
Data Extraction:
Data were extracted on the primary outcome, central venous access device failure, and secondary outcomes: central venous access device complications (central line-associated bloodstream infection, catheter-related bloodstream infection, catheter-related thrombosis, occlusion, catheter removal due to suspected infection, dislodgement, breakage, and local infection). Patient and device data and study details to assess the study quality were also extracted.
Data Synthesis:
A total of 63 studies involving 50,000 central venous access devices (396,951 catheter days) were included. Central venous access device failure was 5% (95% CI, 3-6%), with the highest rates and proportion of failure in hemodialysis catheters. Overall central line-associated bloodstream infection rate was 4.59 per 1,000 catheter days (95% CI, 2.31-6.86), with the highest rate in nontunneled central venous access devices. Removal of central venous access device due to suspected infection was high (17%; 20.4 per 1,000 catheter days; 95% CI, 15.7-25.2).
Conclusions:
Central venous access device complications and device failure is a prevalent and significant problem in the adult ICU, leading to substantial patient harm and increased healthcare costs. The high proportion of central venous access devices removed due to suspicion of infection, despite low overall central line-associated bloodstream infection and catheter-related bloodstream infection rates, indicates a need for robust practice guidelines to inform decision-making surrounding removal of central venous access devices suspected of infection.
More Related Videos
Related Concept Videos
Specialized Care Centers and Settings-II
Rural health centers are specialized care facilities in remote locations with very few medical personnel. The primary care providers who run the centers are mostly Registered Nurse Practitioners. Here, emergency treatment is provided to critically ill or injured patients before they are transferred to the closest hospital. Fortunately, due to advancement in technology, many rural healthcare facilities and professionals have easy access to diagnostic and treatment...
Specialized Care Centers and Settings-I
Daycare centers
They provide several functions. Some facilities care for healthy newborns and children whose parents work, while others are medically focused and care for...
The Central Dogma
Measures of Central Tendency
Critical Region, Critical Values and Significance Level
In hypothesis testing, a sample statistic is converted to a test statistic using z, t, or chi-square distribution. A critical region is an area under the curve in probability distributions demarcated by the critical value. When the test statistic falls in this region, it suggests that the null hypothesis must be rejected. As this region contains all those values of the...
Hemodialysis II: Procedure and Complications