Interscanner and Intrascanner Comparison of Virtual Unenhanced Attenuation Values Derived From Twin Beam Dual-Energy and Dual-Source, Dual-Energy Computed Tomography
Markus M Obmann1, Vanessa Kelsch, Aurelio Cosentino2
1Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA.
Objective:
The aim of the current study was to evaluate the reliability and comparability of virtual unenhanced (VUE) attenuation values derived from scans of a single-source, dual-energy computed tomography using a split-filter (tbDECT) to a dual-source dual-energy CT (dsDECT).
Materials And Methods:
In this retrospective study, comparisons for tbDECT and dsDECT were made within and between different dual-energy platforms. For the interscanner comparison, 126 patients were scanned with both scanners within a time interval of 224 ± 180 days; for the intrascanner comparison, another 90 patients were scanned twice with the same scanner within a time interval of 136 ± 140 days. Virtual unenhanced images were processed off of venous phase series. Attenuation values of 7 different tissues were recorded. Disagreement for VUE HU measurements greater than 10 HU between 2 scans was defined as inadequate.
Results:
The interscanner analysis showed significant difference between tbDE and dsDE VUE CT values (P < 0.01) for 6 of 7 organs. Percentage of cases that had more than 10 HU difference between tbDE and dsDE for an individual patient ranged between 15% (left kidney) and 62% (spleen).The intrascanner analysis showed no significant difference between repeat scans for both tbDECT and dsDECT (P > 0.05). However, intrascanner disagreements for the VUE HU measurements greater than 10 HU were recorded in 10% of patients scanned on the tbDECT and 0% of patients scanned on the dsDECT. The organs with the highest portion of greater than 10 HU errors were the liver and the aorta (both 20%).
Conclusions:
Dual-energy techniques vary in reproducibility of VUE attenuation values. In the current study, tbDECT demonstrated higher variation in VUE HU measurements in comparison to a dsDECT. Virtual unenhanced HU measurements cannot be reliably compared on follow-up CT, if these 2 different dual-energy CT platforms are used.
More Related Videos
09:20Novel Photoacoustic Microscopy and Optical Coherence Tomography Dual-modality Chorioretinal Imaging in Living Rabbit Eyes
Published on: February 8, 2018
07:00Preparing Lamellae from Vitreous Biological Samples Using a Dual-Beam Scanning Electron Microscope for Cryo-Electron Tomography
Published on: August 5, 2021
Related Concept Videos
Free Energy and Equilibrium
Recall that Q is the numerical value of the mass action...
Dual Nature of Electromagnetic (EM) Radiation
Wavelength is the distance between two consecutive peaks (the highest point) or troughs (the lowest point) in the wave. Frequency is the number of...
What is Energy?
Free Energy
Activation Energy
Cell Potential and Free Energy
Thermodynamics is the branch of physics dealing with the relationship between heat and other forms of energy. In an electrochemical cell, chemical energy is converted into electrical energy.
Thus, a link can be predicted between cell potential, free energy change, and the equilibrium constant for the reaction. Cell potential can also be measured as the oxidant or the reducing strength, and similar acid-base strength measures are reflected in equilibrium...