Background:
Taper corrosion in total hip arthroplasty for bearings with metal heads against polyethylene has developed from an anecdotal observation to a clinical problem. Increased taper wear and even gross taper failure have been reported for one particular design. It is hypothesized that corrosion of the female head taper results in taper widening, allowing the cobalt-chromium head to turn on the stem and wear down the softer titanium alloy by abrasive wear, ultimately causing failure. The purpose of this study is to investigate the time course of this process and the general role of taper dimensions and material in this problem.
Methods:
Retrieved cobalt-chromium alloy heads (n = 30, LFIT; Stryker, Mahwah, NJ) and Ti-12Mo-6Zr-2Fe (TMZF) stems (n = 10, Accolade I; Stryker) were available for analysis. Taper material loss was determined using three-dimensional coordinate measurements and scanning. The pristine tip clearance between head and stem was analytically determined. The influence of taper material and taper size on taper deformation and micromotion was investigated using a finite element model.
Results:
Material loss at the head taper increased with time in situ up to a volume of 20.8 mm3 (P < .001). A mean linear material loss above 76 μm at the head taper was analytically confirmed to result in bottoming out, which was observed in 12 heads. The finite element calculations showed significantly larger deformations and micromotions for a small 11/13 TMZF taper combined with a distinctly different micromotion pattern compared to other materials and taper designs.
Conclusion:
A 11/13 TMZF taper design with 36-mm head diameters bears a higher risk for corrosion than larger tapers made from stiffer materials. Failures of this combination are not restricted to the head sizes included in the recall. Patients with this implant combination should be closely monitored.
Related Concept Videos
Muscles that Move the Head
The bilateral sternocleidomastoid, or SCM, and the suprahyoid and infrahyoid muscles are significant head flexors. The SCM muscles originate at the sternum and clavicle and attach to the mastoid process of the temporal bone. The SCM contracts bilaterally to bend the head forward, whereas...
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Corrosion
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...
Gross Anatomy of the Lungs