Error Detection in Intensity-Modulated Radiation Therapy Quality Assurance Using Radiomic Analysis of Gamma
Landon S Wootton1, Matthew J Nyflot2, W Art Chaovalitwongse3
1Department of Radiation Oncology, University of Washington School of Medicine, Seattle, Washington.
Purpose:
To improve the detection of errors in intensity-modulated radiation therapy (IMRT) with a novel method that uses quantitative image features from radiomics to analyze gamma distributions generated during patient specific quality assurance (QA).
Methods And Materials:
One hundred eighty-six IMRT beams from 23 patient treatments were delivered to a phantom and measured with electronic portal imaging device dosimetry. The treatments spanned a range of anatomic sites; half were head and neck treatments, and the other half were drawn from treatments for lung and rectal cancers, sarcoma, and glioblastoma. Planar gamma distributions, or gamma images, were calculated for each beam using the measured dose and calculated doses from the 3-dimensional treatment planning system under various scenarios: a plan without errors and plans with either simulated random or systematic multileaf collimator mispositioning errors. The gamma images were randomly divided into 2 sets: a training set for model development and testing set for validation. Radiomic features were calculated for each gamma image. Error detection models were developed by training logistic regression models on these radiomic features. The models were applied to the testing set to quantify their predictive utility, determined by calculating the area under the curve (AUC) of the receiver operator characteristic curve, and were compared with traditional threshold-based gamma analysis.
Results:
The AUC of the random multileaf collimator mispositioning model on the testing set was 0.761 compared with 0.512 for threshold-based gamma analysis. The AUC for the systematic mispositioning model was 0.717 versus 0.660 for threshold-based gamma analysis. Furthermore, the models could discriminate between the 2 types of errors simulated here, exhibiting AUCs of approximately 0.5 (equivalent to random guessing) when applied to the error they were not designed to detect.
Conclusions:
The feasibility of error detection in patient-specific IMRT QA using radiomic analysis of QA images has been demonstrated. This methodology represents a substantial step forward for IMRT QA with improved sensitivity and specificity over current QA methods and the potential to distinguish between different types of errors.
Related Concept Videos
Quality Assurance
Biological Effects of Radiation
Detection of Gross Error: The Q Test
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Fundamental Attribution Error