MR-Only Brain Radiation Therapy: Dosimetric Evaluation of Synthetic CTs Generated by a Dilated Convolutional Neural
Anna M Dinkla1, Jelmer M Wolterink2, Matteo Maspero3
1Department of Radiation Oncology, University Medical Center Utrecht, Utrecht, The Netherlands.
Purpose:
This work aims to facilitate a fast magnetic resonance (MR)-only workflow for radiation therapy of intracranial tumors. Here, we evaluate whether synthetic computed tomography (sCT) images generated with a dilated convolutional neural network (CNN) enable accurate MR-based dose calculations in the brain.
Methods And Materials:
We conducted a retrospective study of 52 patients with brain tumors who underwent both computed tomography (CT) and MR imaging for radiation therapy treatment planning. To generate the sCTs, a T1-weighted gradient echo MR sequence was selected from the clinical protocol for multiple types of brain tumors. sCTs were created for all 52 patients with a dilated CNN using 2-fold cross validation; in each fold, 26 patients were used for training and the remaining 26 patients were used for evaluation. For each patient, the clinical CT-based treatment plan was recalculated on sCT. We calculated dose differences and gamma pass rates between CT- and sCT-based plans inside body and planning target volume. Geometric fidelity of the sCT and differences in beam depth and equivalent path length were assessed between both treatment plans.
Results:
sCT generation took 1 minute per patient. Over the patient population, the mean absolute error of the sCT within the intersection of body contours was 67 ± 11 HU (±1 standard deviation [SD], range: 51-117 HU), and the mean error was 13 ± 9 HU (±1 SD, range: -2 to 38 HU). Dosimetric analysis showed mean deviations of 0.00% ± 0.02% (±1 SD, range: -0.05 to 0.03) for dose within the body contours and -0.13% ± 0.39% (±1 SD, range: -1.43 to 0.80) inside the planning target volume. Mean γ1mm/1% was 98.8% ± 2.2% for doses >50% of the prescribed dose.
Conclusions:
The presented dilated CNN generated sCTs from conventional MR images without adding scan time to the acquisition. Dosimetric evaluation suggests that dose calculations performed on the sCTs are accurate and can therefore be used for MR-only intracranial radiation therapy treatment planning.
Related Concept Videos
Generating Electromagnetic Radiations
Convolution Properties II
The width property indicates that if the durations of input signals are T1 and T2, then the width of the output response equals the sum of both durations, irrespective of the shapes of the two functions. For instance, convolving two rectangular pulses with durations of 2 seconds and 1 second results in a function with a width of 3 seconds.
The area property asserts that the area under the...
Biological Effects of Radiation
Convolution Properties I
The commutative property reveals that the input and the impulse response of an LTI (Linear Time-Invariant) system can be interchanged without affecting the output:
Cardiomyopathy II: Dilated Cardiomyopathy
Protein Networks
These interactions can be represented through maps depicting protein-protein interaction networks, represented as nodes and edges. Nodes are circles that are representative of a protein,...