Related Experiment Video
Updated: Feb 6, 2026

Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
Ultrasensitive quantification of cardiac troponin I by a Single Molecule Counting method: analytical validation and
Alvaro Garcia-Osuna1, David Gaze2, Margarita Grau-Agramunt1
1Department of Clinical Biochemistry, IIB-Hospital de Sant Pau, Barcelona, Spain.
Aim:
To evaluate analytical and biological characteristics of the Singulex Clarity® cTnI assay, based upon Single Molecule Counting technology.
Methods:
Assay's analytical sensitivity, precision, linearity, hook effect, cross-reactivity or interference by endogenous and exogenous substances, stability, 99th reference percentile [p99th] in EDTA plasma were evaluated in single or multi-site studies.
Results:
Detection limit was 0.12 ng/L. Sensitivity was 0.14 ng/L at 20% CV (functional sensitivity) and 0.53 ng/L at 10% CV. Imprecision was 3.16%-10.0% in a multi-lot, single-site study, and 5.5%-12.0% in a single-lot, multi-site study; assay was linear from 0.08 to 25,000 ng/L. No hook effect was observed; any cross-reactivity/interference exceeded the 10%. Healthy subjects were recruited using clinical history, normal NT-proBNP and eGFR (n = 560) or plasma creatinine (n = 535) as inclusion criteria. cTnI was detectable in 96.8% of healthy subjects. The p99th were 8.01 (eGFR used) and 8.15 ng/L (plasma creatinine); both were measured with ≤5.7% CV. Median cTnI were significantly higher in older and male than in young and female subjects.
Conclusions:
The Singulex Clarity cTnI assay show analytical features and % detection in healthy subjects that improve the corresponding values of most of the existing high-sensitivity cTnI assays.
Related Concept Videos
Development of Analytical Methods
Biological Methods for Microbial Control
Reliability and Validity
What is Conservation Biology?
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...