Safer employment of nitrous oxide in anesthesia machines-a technical simulation
Thomas Mitterlechner1, Peter Paal2, Daniel C Schroeder3
1Department of Anesthesiology, Privatklinik Hochrum, Sanatorium der Kreuzschwestern, Rum, Austria.
Abstract:
Several incidents of anesthesia-attributed mortality in the past were caused by misconnection of gas pipelines resulting in ventilation with pure nitrous oxide. A simple safety feature may be to "mark" nitrous oxide with a lower pressure than oxygen and room air within the hospital's gas pipeline system. Then, any misconnection of gas pipelines could be detected by pressure differences with a manometer in the anesthesia machine. To check technical suitability, we tested maximum achievable nitrous oxide flows of an anesthesia machine at different pressures in the nitrous oxide supply line. Using decreased pressures for nitrous oxide compared to oxygen did not result in decreased nitrous oxide flows, as long as pressure in the nitrous oxide supply line was >1500 hPa. A concept of different pressures for nitrous oxide and oxygen could be used to technically differentiate between those two gases, and to avoid potentially fatal misconnections.
Related Concept Videos
Oxidation Numbers
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Stages of General Anesthesia
Machines
A free-body diagram of the...
Pyruvate Oxidation
First, the enzyme pyruvate dehydrogenase removes the carboxyl group from pyruvate and releases it as carbon dioxide. The stripped molecule is then oxidized and releases electrons, which are then picked up by NAD+...
Machines: Problem Solving II