Medical physics 3.0 versus 1.0: A case study in digital radiography quality control
Diana E Carver1, Charles E Willis1, Paul J Stauduhar1
1Department of Imaging Physics, University of Texas M. D. Anderson Cancer Center, Houston, TX, USA.
Purpose:
The study illustrates how a renewed approach to medical physics, Medical Physics 3.0 (MP3.0), can identify performance decrement of digital radiography (DR) systems when conventional Medical Physics 1.0 (MP1.0) methods fail.
Methods:
MP1.0 tests included traditional annual tests plus the manufacturer's automated Quality Assurance Procedures (QAP) of a DR system before and after a radiologist's image quality (IQ) complaint repeated after service intervention. Further analysis was conducted using nontraditional MP3.0 tests including longitudinal review of QAP results from a 15-yr database, exposure-dependent signal-to-noise (SNR2 ), clinical IQ, and correlation with the institutional service database. Clinical images were analyzed in terms of IQ metrics by the Duke University Clinical Imaging Physics Group using previously validated software.
Results:
Traditional metrics did not indicate discrepant system performance at any time. QAP reported a decrease in contrast-to-noise ratio (CNR) after detector replacement, but remained above the manufacturer's action limit. Clinical images showed increased lung noise (Ln), mediastinum noise (Mn), and subdiaphragm-lung contrast (SLc), and decreased lung gray level (Lgl) following detector replacement. After detector recalibration, QAP CNR improved, but did not return to previous levels. Lgl and SLc no longer significantly differed from before detector recalibration; however, Ln and Mn remained significantly different. Exposure-dependent SNR2 documented the detector operating within acceptable limits 9 yr previously but subsequently becoming miscalibrated sometime before four prior annual tests. Service records revealed catastrophic failure of the computer containing the original detector calibration from 11 yr prior. It is likely that the incorrect calibration backup file was uploaded at that time.
Conclusions:
MP1.0 tests failed to detect substandard system performance, but MP3.0 methods determined the root cause of the problem. MP3.0 exploits the wealth of data with more sensitive performance indicators. Data analytics are powerful tools whose proper application could facilitate early intervention in degraded system performance.
More Related Videos
Related Concept Videos
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Physical Methods for Controlling Microbial Growth: Temperature
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Physical and Chemical Properties of Matter
Inhaled Medications