Related Experiment Video
Updated: Feb 6, 2026

A Rat Model of Ventricular Fibrillation and Resuscitation by Conventional Closed-chest Technique
Published on: April 26, 2015
A pilot study evaluating adaptive closed-loop fluid resuscitation during states of absolute and relative hypovolemia
Behnood Gholami1, Wassim M Haddad2, James M Bailey3
1Autonomous Healthcare Inc, Hoboken, NJ, USA.
Objective:
To evaluate and determine the performance of a partially automated as well as a fully automated closed-loop fluid resuscitation system during states of absolute and relative hypovolemia.
Design:
Prospective experimental trial.
Setting:
Research laboratory.
Animals:
Five adult Beagle dogs.
Methods:
Isoflurane anesthetized mechanically ventilated dogs were subjected to absolute hypovolemia (controlled: 2 trials; uncontrolled: 3 trials), relative hypovolemia (2 trials), and the combination of relative and absolute controlled hypovolemia (2 trials). Controlled and uncontrolled hypovolemia were produced by withdrawing blood from the carotid or femoral artery. Relative hypovolemia was produced by increasing the isoflurane concentration (1 trial) or by infusion of intravenous sodium nitroprusside (1 trial). Relative hypovolemia combined with controlled absolute hypovolemia was produced by increasing the isoflurane concentration (1 trial) and infusion of IV sodium nitroprusside (1 trial). Hemodynamic parameters including stroke volume variation (SVV) were continuously monitored and recorded in all dogs. A proprietary closed-loop fluid administration system based on fluid distribution and compartmental dynamical systems administered a continuous infusion of lactated Ringers solution in order to restore and maintain SVV to a predetermined target value.
Measurements And Main Results:
A total of 9 experiments were performed on 5 dogs. Hemodynamic parameters deteriorated and SVV increased during controlled or uncontrolled hypovolemia, relative hypovolemia, and during relative hypovolemia combined with controlled hypovolemia. Stroke volume variation was restored to baseline values during closed-loop fluid infusion.
Conclusions:
Closed-loop fluid administration based on IV fluid distribution and compartmental dynamical systems can be used to provide goal directed fluid therapy during absolute or relative hypovolemia in mechanically ventilated isoflurane anesthetized dogs.
Related Concept Videos
Bioavailability Study Design: Absolute Versus Relative Bioavailability
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Pilot and Numeric Relaying
Absolute Value Inequalities
Mean Absolute Deviation
Let us consider a dataset containing the number of unsold cupcakes in five shops: 10, 15, 8, 7, and 10. Initially, calculate the sample mean. Then calculate the deviation, or the difference, between each data value and the mean. Next, the absolute values of these deviations are added and divided by the sample size to...
Cardiopulmonary Resuscitation I: Adult