Related Experiment Video
Updated: Feb 6, 2026

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Ultrasound Evaluation of Puncture Sites After Deployment of Two Different Types of Vascular Closure Devices: A
Young Jin Heo1, Hae Woong Jeong2, Jin Wook Baek1
1Department of Radiology, Inje University Busan Paik Hospital, Bokji-ro 75, Busanjin-Gu, Busan, Korea, 47392.
Purpose:
The aim of this study was to compare the ultrasonographic findings of femoral puncture sites and the complications of Perclose ProGlide® and FemoSeal™ after neurointerventional procedures.
Methods:
In this prospective, single-center study, we randomly assigned 155 femoral puncture sites to treatment with Perclose ProGlide® or FemoSeal™. We hypothesized that the two different types of VCD cause different vascular changes. Ultrasonography of the femoral puncture sites was performed 24 h after the procedure and at an outpatient visit after 6 months. The intima-media thickness (IMT), vessel diameter, and minimal luminal diameter of the common femoral artery were measured; the perivascular soft tissue change and absorption of the hemostatic material were observed. The device failure rate and vascular complications associated with each device were also evaluated.
Results:
Follow-up ultrasonography was performed at a median follow-up time of 187.0 days (range 147-240 days) after the initial ultrasonography. The IMT on follow-up ultrasonography was significantly higher in patients who received FemoSeal™ (P = 0.0000). Intimal hyperplasia and partial absorption of the hemostatic material were significantly more frequent in patients who received FemoSeal™. The vessel diameters on initial and follow-up ultrasonography were not significantly different, but the minimal diameter on follow-up ultrasonography was significantly lower in patients who received FemoSeal™. Device failure and pseudoaneurysms occurred at 9 and 3 puncture sites in patients who received Perclose ProGlide®, respectively.
Conclusions:
Intimal hyperplasia was more frequently observed in patients who received FemoSeal™. However, more device failures and pseudoaneurysms occurred in patients who received Perclose ProGlide®.
Level Of Evidence:
Step 2 (level 2).
More Related Videos
Related Concept Videos
Conserved Binding Sites
Binding sites are often located in large pockets, and if their location on a protein’s surface is unknown, it can be predicted using various approaches. The energetic method computationally...
Conserved Binding Sites
Ligand Binding Sites
Protein-ligand interactions are quite specific; even though numerous potential ligands surround a cellular protein at any given time, only a particular ligand can bind to that protein. Moreover, a ligand binds only to a dedicated area on the surface of the protein, known as the...
Ligand Binding Sites
Imaging Studies VII: Vascular Imaging
Seedless Vascular Plants