Initial single centre experience with the novel Rhythmia© high density mapping system in an all comer collective of
Korbinian Lackermair1, Stefanie Kellner1, Antonia Kellnar1
1Department of Medicine I, University Hospital Munich, Ludwig Maximilians University, Munich, Germany.
Background:
A novel, automatically annotating ultra-high density mapping system (Rhythmia©, Boston Scientific) collects a high number and quality of electrograms (EGMs). So far, data on general use in the electrophysiological laboratory are sparse.
Methods:
We retrospectively analyzed all our ablations using Rhythmia and recorded patient clinical data, procedural parameters, and mapping parameters including the count of EGMs, mapping time, and mapping volume. Where appropriate, procedural parameters were compared over time to assess a learning curve.
Results:
400 patients underwent ablation of atrial fibrillation (n = 202), typical (n = 16) or atypical atrial flutter (n = 49), VT (n = 48), PVC (n = 35), accessory pathways (n = 14), AVNRT (n = 4), and focal atrial tachycardia (n = 32). System use was feasible, as no procedure had to be stopped for technical reasons and no ablation had to be withheld because of mapping failure, and safe, with an overall complication rate of 2.25%. Initial restrictions in manoeuvrability of the mapping catheter were overcome rapidly, as indicated by a significant decrease of fluoroscopy time (20 vs. 14 min, p = 0.02), use of contrast agent (50 vs. 40 ml; p < 0.01), and (not significant) lower procedure times (194 vs. 170 min; p = 0.12; comparing the first with the last third of patients undergoing pulmonary vein isolation only procedure). Ablation of complex left atrial, focal and ventricular tachycardias benefited from the reliable automatic annotation of a high number of EGMs.
Conclusion:
The use of the Rhythmia is feasible and safe. Initial restrictions in manoeuvrability of the Orion mapping catheter were overcome rapidly. The procedures that benefit the most from ultra-high density mapping are complex left atrial tachycardias, focal tachycardias, and ventricular tachycardias.
Related Concept Videos
Data Collection by Experiments
An example of the experimental method is a public...
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Transcription Initiation
The promoters and enhancers and their accessory proteins allow tight regulation of...
What is an Experiment?
Density