Related Experiment Video
Updated: Feb 6, 2026

In Situ Surface Temperature Measurement in a Conveyor Belt Furnace via Inline Infrared Thermography
Published on: May 30, 2020
Comparison of Non-contact Infrared Temperature with Digital Axillary Temperature
Varghese Abraham1, Elizabeth Johnson2, Khristina Deep2
1Christian Hospital, Chhatarpur, Madhya Pradesh, India. varghesenettaabraham@gmail.com.
Abstract:
We measured the Non-contact infrared temperature (NCIT) from the forehead, chest and abdomen, and compared it with Digital axillary temperature (DAT) by Bland Altman Plot. The DAT agreed better with NCIT chest (mean difference 0.13, 95% limit of agreement 0.08, 0.18) as compared to NCIT forehead and abdomen.
More Related Videos
Related Concept Videos
Body Temperature
Body Temperature
The average body temperature is approximately 37°C (98.6°F) and typically ranges from 36.1–37.2°C...
Effects of Temperature on Free Energy
Factors Affecting Body Temperature
Factors may include:
Increased Body Temperature
Decreased Body Temperature