Endotracheal Intubation of Rabbits Using a Polypropylene Guide Catheter
Published on: November 13, 2017
A Comparison of Laser-Protected Endotracheal Tubes
James A Burns1, Stephen D Adlard1, James B Kobler1
11 Center for Laryngeal Surgery and Voice Rehabilitation, Department of Surgery, Harvard Medical School, Massachusetts General Hospital, Boston, Massachusetts, USA.
Objectives:
To compare the physical characteristics of 3 laser-protected endotracheal tubes (LPETs) commonly used in endoscopic laser surgery. To report potential intraoperative problems related to LPET use and suggest practical solutions.
Study Design:
Comparative analysis.
Setting:
Academic laboratory.
Subjects And Methods:
Physical characteristics of the Mallinckrodt Laser-Flex (MTL), Medtronic Laser-Shield II (ML-II), and Rusch LaserTubus (RL) were compared. The effect of bending LPETs on airflow resistance was estimated with a pressure transducer. The force required to pull each tube through the glottis and the pressure exerted during this maneuver were measured in a fresh cadaveric human larynx.
Results:
The design features and physical characteristics of LPETs differ, including varying balloon-tip lengths. Bending LPETs to acute angles caused significant pressure increase within the RL tube (Δ 3.42 cm H2O) and minimal change within the ML-II (Δ 0.12 cm H2O) and MTL (Δ 0.21 cm H2O) tubes. The average force required to pull the RL (48.12 g, P = .003) and MTL (282.4 g, P = .001) tubes through the glottis was 7.6× and 44.5× greater than that for the ML-II (6.39 g). When pulled through the vocal folds, the ML-II cuff exerted no detectable pressure, whereas higher pressures were measured for the RL (2.2 cm H2O) and MTL (6.5 cm H2O) tubes.
Conclusion:
The ML-II tube had the most favorable characteristics, with minimal pressure during extubation and resistance to kinking. The RL tube kinks readily with a resultant increase in resistance to airflow. The MTL tube extends farther into the trachea due to a relatively elongated balloon-tip configuration. Future LPET designs should incorporate features that avoid intraoperative difficulties related to airway protection and ventilation.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Protection of Alcohols
Protection
It defines a protecting group as the masking agent to make the more reactive species inert to a given set of conditions. This concept is depicted via the illustration of liquid flow through different outlets in an assembly of pipes. The analogy helps to understand the role...
Zones of Protection
Protective zones are defined by closed dashed lines, containing one or more components. A key characteristic of these zones is the strategic placement of...