Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Influence of calibration on digital templating of hip arthroplasty
Christoph Kolja Boese1, Sebastian Wilhelm2, Stefan Haneder3
1Department of Orthopaedic and Trauma Surgery, University Hospital of Cologne, Joseph-Stelzmann-Str. 9, 50924, Cologne, Germany. christoph.boese@uk-koeln.de.
Introduction:
Digital templating for total joint replacement is the current standard. For image calibration, external calibration markers (ECM) are used. However, there are concerns regarding the precision of the method. This study aimed to identify the direct influence of calibration errors on digital templating.
Patients And Methods:
A retrospective analysis of 100 post-operative radiographs with unilateral total hip arthroplasty was performed. The magnification factor of the ECM and of the internal prosthetic femoral head (ICM) as a reference value was calculated for each radiograph. Two blinded observers performed templating of the contralateral hip using a randomized list for all radiographs and both markers. The component size templated by the ECM magnification was compared to the reference by the ICM magnification.
Results:
Mean magnification factors of ICM and ECM differed significantly (p = 0.006). The absolute difference was 5.2% (range 0.0-23.3%, SD 4.8%). Templating of the acetabular or the femoral component showed no significant differences (p = 0.120, p = 0.599). Differences of more than one size were found in 26% of the acetabular components and 14% of the femoral components and differences over two sizes in 10% respectively 3%. Correlation coefficients for magnification error and size differences of acetabular components were - 0.645 (p < 0.001) and for the femoral component - 0.607 (p < 0.001).
Interpretation:
The calibration error of external calibration markers in digital templating for hip replacement influences component sizes significantly. Thus, correct positioning of ECM is of utmost importance.
Related Concept Videos
DNA as a Genetic Template
DNA as a Genetic Template
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Correlation Coefficient
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...