Related Experiment Video
Updated: Feb 6, 2026

Preparation of Aligned Steel Fiber Reinforced Cementitious Composite and Its Flexural Behavior
Published on: June 27, 2018
Cyclic flexural fatigue resistance of NiTi Controlled Memory and Blue Technology instruments after torsional
Túlio César Modesto1, Eufemia Carolina Peláez Acosta1, Pedro Damas Resende2
1Departmento de Odontologia Restauradora, Faculdade de Odontologia, Universidade Federal de Minas Gerais, Belo Horizonte, MG, Brasil.
Abstract:
Objective The aim of this study was to evaluate the influence of torsional preloading on the cyclic flexural fatigue resistance of thermally treated NiTi instruments. Material and Methods Ten new instruments New Hyflex CM (HF 30.06; Coltene/Whaladent Inc.), Typhoon CM (TYP 30.06; Clinician's Choice Dental Products) and Vortex Blue (VB 30.06; Dentsply Tulsa Dental) were chosen, based on geometry and specific characteristics of the manufacturing process. The new instruments of each system were tested in a bench device to determine their fatigue resistance through mean value of number of cycles to failure (Nf) (Control Group - CG). Another group of 10 new HF, TYP and VB instruments were submitted to 20 cycles of torsional straining between 0° and 180° (Experimental Group - EG) and then submitted to fatigue until rupture under the same conditions of the CG. Tested instruments were examined by scanning electron microscopy (SEM). Data were analyzed using one-way analysis of variance and post hoc Tukey's test (α=.05). Results Higher fatigue resistance was accomplished by HF instruments, followed by VB and TYP (p<0.05). During the torsional preloading, the lowest mean torque value was observed for TYP instruments (p<0.05). The torsional preload caused a significant reduction in the Nf values (p<0.05) of about 20%, 39% and 45% for instruments HF, VB and TYP, respectively. Longitudinal cracks, generated during the torsional preloading, were present in VB files, but were not observed in the CM instruments (HF and TYP). Conclusions In conclusion, the flexural fatigue resistance of thermally treated instruments is diminished after cyclic torsional loading.
Related Concept Videos
Torsion of Noncircular Members
Torsional Pendulum
As long as the rigid body's angular displacement is small, its oscillation can be modeled as a linear angular oscillation. The amplitude of the oscillation is an angle. The role of mass is played...
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to its distance...
Fatigue
Bending and Torsional Moments
The reaction developed in a structural element when subjected to an external force causes the element to bend. When a structural element bends upwards, it creates compressive normal forces on the top and tensile normal forces on the bottom, resulting in a couple that determines the bending...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...