Technical Note: Real-time image-guided adaptive radiotherapy of a rigid target for a prototype fixed beam
P Z Y Liu1, D T Nguyen1, I Feain2
1ACRF Image X Institute, University of Sydney Central Clinical School, Sydney, NSW, Australia.
Medical Physics
|August 23, 2018
Summary
Fixed beam radiotherapy with real-time adaptation achieves clinical accuracy comparable to conventional rotating beam systems, even with target motion. This simplifies linear accelerator design while maintaining treatment precision.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- Fixed beam radiotherapy simplifies linear accelerator design by using couch movement instead of gantry rotation.
- Maintaining clinical accuracy in fixed beam systems requires addressing potential target motion.
Purpose of the Study:
- To investigate the feasibility of delivering accurate fixed beam radiotherapy treatments.
- To compare the accuracy of fixed beam vs. conventional rotating beam radiotherapy with real-time image guidance.
- To assess the impact of rigid target motion on treatment accuracy.
Main Methods:
- A prototype fixed beam system with a vertical linac and rotational stage was developed.
- Kilovoltage Intrafraction Monitoring (KIM) and real-time multi-leaf collimator (MLC) tracking were employed for adaptive radiotherapy.
- Film dosimetry was used to compare dose distributions between fixed and rotating beam geometries with and without real-time adaptation.
Main Results:
- Dose distributions were equivalent between fixed and rotating beam geometries when using KIM and MLC tracking for adaptation.
- Gamma analysis demonstrated >98% agreement (2%/2 mm criteria) with adaptation for static and periodic motion.
- Real-time adaptation maintained accuracy despite rigid target motion.
Conclusions:
- Fixed beam radiotherapy with real-time adaptation is dosimetrically equivalent to conventional rotating beam treatments.
- This approach maintains high clinical accuracy for rigid targets, even with motion.
- Simpler fixed beam geometry is viable for achieving accurate real-time adaptive radiotherapy.
Related Concept Videos
Concepts and Prototypes
548
The human nervous system handles vast amounts of information by translating sensory stimuli into neural impulses, which the brain processes, creating thoughts expressed through language or stored as memories. The brain also synthesizes information from emotions and memories, which significantly influence thoughts and behaviors. This intricate process creates a comprehensive mental picture.
The brain organizes this information using concepts, which are mental categories grouping linguistic data,...
The brain organizes this information using concepts, which are mental categories grouping linguistic data,...
548
Real Time RT-PCR
65.3K
Real-time reverse transcription-polymerase chain reaction, or Real-time RT-PCR, is an analytical tool used to determine the expression level of target genes. The method involves converting mRNA to complementary DNA with the help of an enzyme known as reverse transcriptase, followed by the PCR amplification of the cDNA. These two processes can be performed simultaneously in a single tube or separately as a two-step reaction.
The real-time quantification of the number of amplified products is...
The real-time quantification of the number of amplified products is...
65.3K
Beams
1.9K
Beams are integral components of structural engineering and construction, designed to support loads applied at various points along their length. These long, straight members can be classified based on geometry, cross-section, support type, and equilibrium condition.
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
1.9K
Rigid Body Equilibrium Problems - I
5.5K
A rigid body is said to be in static equilibrium when the net force and the net torque acting on the system is equal to zero. To solve for rigid body equilibrium problems, do the following steps.
5.5K
Rigid Body Equilibrium Problems - II
8.0K
A rigid body is in static equilibrium when the net force and the net torque acting on the system are equal to zero.
Consider two children sitting on a seesaw, which has negligible mass. The first child has a mass (m1) of 26 kg and sits at point A, which is 1.6 meters (r1) from the pivot point B; the second child has a mass (m2) of 32 kg and sits at point C. How far from the pivot point B should the second child sit (r2) to balance the seesaw?
Consider two children sitting on a seesaw, which has negligible mass. The first child has a mass (m1) of 26 kg and sits at point A, which is 1.6 meters (r1) from the pivot point B; the second child has a mass (m2) of 32 kg and sits at point C. How far from the pivot point B should the second child sit (r2) to balance the seesaw?
8.0K
Fixed Action Patterns
17.7K
A fixed action pattern (FAP) is a specific, hard-wired sequence of behaviors that occurs in response to an external stimulus, called a sign stimulus. The behavior is “fixed” because it is essentially unchangeable—proceeding similarly across individuals of a species every time it occurs.
17.7K