Toward a novel dosimetry system using acrylic disk radiation sensor for proton pencil beam scanning
Shinhaeng Cho1, Nuri Lee2, Sanghyeon Song3
1Proton Therapy Center, National Cancer Center, Goyang, Korea.
Purpose:
Fabricate an acrylic disk radiation sensor (ADRS) and characterize the photoluminescence signal generated from the optical device as basis for the development and evaluation of a new dosimetry system for pencil beam proton therapy.
Methods:
Based on the characteristics of the proposed optical dosimetry sensor, we established the relation between the photoluminescence output and the applied dose using an ionization chamber. Then, we obtained the relative integral depth dose profiles using the photoluminescence signal generated by pencil beam irradiation at energies of 99.9 and 162.1 MeV, and compared the results with the curve measured using a Bragg peak ionization chamber.
Results:
The relation between the photoluminescence output and applied dose was linear. In addition, the ADRS was dose independent for beam currents up to 6 Gy/min, and the calibration factor for energy was close to 1. Hence, the energy dependence on the optical device can be disregarded. The integral depth dose profiles obtained for the ADRS suitable agreed with the curve measured in the Bragg peak ionization chamber without requiring correction.
Conclusions:
These results suggest that the ADRS is suitable for dosimetry measurements in pencil beam scanning, and it will be employed as a low-cost and versatile dosimetry sensor in upcoming developments.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Faraday Disk Dynamo
Leaky Scanning
Radiation: Applications
The average...
Electric Field of a Charged Disk
The system's symmetry is in the cylindrical directions across the plane of the charge. As a result, the electric fields created by various surface charge elements nullify each other in the direction parallel to the surface. Thereby, the resulting electric field is perpendicular to the plane. Since the disk is...