Characterisation of two new radiochromic gel dosimeters TruView™ and ClearView™ in combination with the vista™
J Colnot1, C Huet1, R Gschwind2
1Institut de Radioprotection et de Sûreté Nucléaire (IRSN), Service de Recherche en Dosimétrie, Laboratoire de Dosimétrie des Rayonnements Ionisants, Fontenay-aux-Roses, France.
Purpose:
This study aims at characterising the properties of TruView™ and ClearView™ two new gel dosimeters (Modus Medical Devices Inc.) and at studying the feasibility of relative dosimetry using these dosimeters and the Vista™ Optical CT scanner to accurately evaluate dose.
Methods:
In this work, we investigated key dosimetric aspects (dose response, energy and dose rate dependence) and stability of these radiochromic gels initiated in preliminary works (Huet et al., 2017; Colnot et al., 2017) using spectrophotometric measurements. Moreover, by mean of optical CT scanning (Vista™), their performances to measure relative depth dose (PDD) and cross profiles were analysed.
Results:
TruView™ and ClearView™ present a linear dose response up to 20 Gy and up to 80 Gy respectively, independent of both photon beam energy (4-18 MV) and dose rate (up to 9.9 Gy/min) (Huet et al., 2017; Colnot et al., 2017). ClearView™ response proves to be stable for a week post-irradiation and uniform within the batch whereas TruView™ presents an unstable but uniform response. Optical CT scanning generates errors due to stray light that need to be corrected in order to use these gels; ClearView™ scanning particularly requires important precautions. After corrections, those gels used in combination with the Vista™ scanner show promising spatial and dosimetric precision (dose difference <5%). Finally, TruView™ is reusable and presents excellent reproducible response (maximum 3% difference) and the ClearView™ dosimeter presents good spatial stability (0.5% difference after 6 days).
Conclusion:
This study provides important knowledge about two gel dosimeters presenting interesting dosimetric properties. A study is ongoing to benchmark those promising candidates for clinical dose verification.
More Related Videos
06:43Effective Analysis of Human Exposure Conditions with Body-worn Dosimeters in the 2.4 GHz Band
Published on: May 2, 2018
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such...
DNA Agarose Gel Electrophoresis
Gel extraction follows five major steps: running gel electrophoresis to separate fragments, isolating the individual bands, extracting DNA from those bands, and removing the dye and salts from the extracted mixture to obtain pure DNA.
In cloning experiments, both the insert and vector DNA...