Related Experiment Video
Updated: Feb 6, 2026

Oral Biofilm Sampling for Microbiome Analysis in Healthy Children
Published on: December 31, 2017
Comparison among Ocular Response Analyzer, Corvis ST and Goldmann applanation tonometry in healthy children
Ramin Salouti1, Ali Agha Alishiri2, Reza Gharebaghi1
1Department of Ophthalmology, Poostchi Ophthalmology Research Center, Shiraz University of Medical Sciences, Shiraz 7134814336, Iran.
Aim:
To explore the relationship between different parameters of Ocular Response Analyzer (ORA) and Corvis ST (CST) in a sample of healthy Iranian school-aged children and the relationship between parameters of these 2 instruments against intraocular pressure (IOP), measured by the Goldmann applanation tonometer (GAT-IOP), age and gender, and find possible correlation between ORA and CST with GAT.
Methods:
This cross-sectional study included 90 healthy children. A general interview and complete eye examination were performed. Following successful GAT-IOP measurement, ORA and CST were conducted. The CST parameters were A 1/2 length (A1L, A2L), A 1/2 velocity (A1V, A2V), highest concavity deformation amplitude (HCDA), radius of curvature (RoC), peak distance (PD), central corneal thickness (CCT) and IOP. The ORA parameters were corneal hysteresis (CH), corneal resistance factor (CRF), Goldmann-correlated IOP (IOP-G) and corneal compensated IOP (IOP-CC). Extracted data was analyzed using the Statistical Package for Social Science software.
Results:
Totally 39 males with age of 9.08±1.60 (6-12)y and 51 females with age of 8.96±1.55 (6-13)y were included. Many CST parameters were significantly correlated with CH, CRF, IOP-G and IOP-CC. Some CST parameters had a significant correlation with GAT-IOP, including IOP-CST in both eyes and HCDA, A2L, PD, and RoC in the left eye, but none with age, except A2L in the right eye. The CRF measurement showed a significant correlation with GAT-IOP in both eyes and CH in the right eye, yet, none with age. Among all CST and ORA parameters, CCT-CST in both eyes and A1L in right eye had a significant correlation with gender, although this was a negligible negative correlation. Comparison of mean IOP values by different devices showed a significantly highest IOP overestimation by CST and lowest by IOP-CC compared with GAT. Also, IOP-G versus IOP-CST significantly had the lowest IOP overestimation among others. Overall, either low positive correlation or negligible correlation was found between IOP measurements by 3 instruments.
Conclusion:
The study finds the highest IOP overestimation by CST and lowest by IOP-CC compared with GAT. Overall, either low positive correlation or negligible correlation is found between IOP measurements by the 3 instruments.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Drug Dosing: Infants and Children

