Design and evaluation of a new bioelectrical impedance sensor for micro-surgery: application to retinal vein
Laurent Schoevaerdts1, Laure Esteveny2, Andy Gijbels2
1Department of Mechanical Engineering, KU Leuven - University of Leuven, Louvain, Belgium. laurent.schoevaerdts@kuleuven.be.
Purpose:
Nowadays, millions of people suffer from retinal vein occlusion, a blind-making eye disease. No curative treatment currently exists for this vascular disorder. However, a promising treatment consists in injecting a thrombolytic drug directly inside the affected retinal vessel. Successfully puncturing miniature vessels with diameters between 50 and 400 [Formula: see text] remains a real challenge, amongst others due to human hand tremor, poor visualisation and depth perception. As a consequence, there is a significant risk of double-puncturing the targeted vessel. Sub-surfacic injection of thrombolytic agent could potentially lead to severe retinal damage.
Methods:
A new bio-impedance sensor has been developed to visually display the instant of vessel puncture. The physical working principle of the sensor has been analysed, and a representative electrical model has been derived. Based on this model, the main design parameters were derived to maximise the sensor sensitivity. A detailed characterisation and experimental validation of this concept were conducted.
Results:
Stable, repeatable and robust impedance measurements were obtained. In an experimental campaign, 35 puncture attempts on ex vivo pig eyes vessels were conducted. A confusion matrix shows a detection accuracy of 80% if there is a puncture, a double puncture or no puncture. The 20% of inaccuracy most probably comes from the limitations of the employed eye model and the experimental conditions.
Conclusions:
The developed bio-impedance sensor has shown great promise to help in avoiding double punctures when cannulating retinal veins. Compared to other puncture detection methods, the proposed sensor is simple and therefore potentially more affordable. Future research will include validation in an in vivo situation involving vitreoretinal surgeons.
More Related Videos
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
07:44Author Spotlight: Implementation of BIVA for Analyzing Disease Risk Factors in Patients with Low Body Cell Mass
Published on: July 14, 2023
Related Concept Videos
Design Example: Application of Archimedes' Principle
The volume of seawater displaced by the block is determined by first calculating the block's weight. This is done by multiplying the...
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Factorial Design
Group Design
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...