A comparison of intraoral spectrophotometers-Are there user-specific differences?
Sam Lennert Blum1, Matthias Horn2, Constanze Olms1
1Department of Dental Prosthodontics and Materials Science, University of Leipzig, Liebigstraße 12, Leipzig, 04103, Germany.
Objective:
The objective of this study was to determine the potential impact of a positioning splint on the determination of the correct tooth color using VITA Easyshade 4 (VE4) or VITA Easyshade 5 (VE5) (VITA Zahnfabrik, Germany).
Materials And Methods:
Dental students (n = 50) in their third year and one expert determined the tooth shade of an extracted human tooth using VE4 and VE5, in presence or absence of a positioning splint. The determined shade values (L*a*b*) were collected and the delta E was calculated. A statistical evaluation was performed with the program SPSS® (P < .05).
Results:
There was no significant difference in the determination of the correct tooth color (VITA 3D-Master shade), independent from either the expert or the students positioning splint, and independent of using VE4 (P = 1; P = .092) or VE5 (P = .125; P = .25).
Conclusion:
Using VE4 or VE5, a positioning splint has no effect on the determination of the correct tooth color (VITA 3D-Master shade). For an exact measurement of the L*a*b values in clinical studies, a positioning splint should be used. A person untrained in working with the devices can determine the correct tooth shade more frequently with VE5 than with VE4.© 2018 Wiley Periodicals, Inc.
Clinical Significance:
Correctly determining the tooth shade is an essential treatment step in esthetic reconstructive dentistry. Color is probably one of the most important determinants of esthetics in dentistry. Factors such as lighting conditions, gender, age, experience, and color vision impairment affect the process of shade matching. These deficits brought about the desire for an objective and reproducible shade determination method with digital devices for shade determination. One of the most proven and frequently used digital shade determination devices is the "VITA Easyshade" (VITA Zahnfabrik, Bad Säckingen, Germany). At this point in time, there are no (published) reports on the measurement accuracy of the new VE 5. In addition, there are, as of yet, no investigations comparing the VE 4 and VE 5.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Electric Potential and Potential Difference
When a test charge moves from the initial to the final position, the electric potential difference between those positions is defined as the ratio of the change in the potential energy to the charge on the...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Identifying Statistically Significant Differences: The F-Test
Sum and Difference OpAmps
A summing amplifier, or an adder, utilizes an op-amp to merge multiple input signals into a single output signal. When audio signals are introduced into its input channels, the input resistors initiate currents that traverse feedback resistors, resulting in an output voltage. Applying Kirchhoff's current...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...