Body Weight-Based Protocols During Whole Body FDG PET/CT Significantly Reduces Radiation Dose without Compromising
Charbel Saade1, Ahmad Ammous1, Alain S Abi-Ghanem1
1Diagnostic Radiology Department, American University of Beirut Medical Center, Lebanon. P.O.Box: 11-0236 Riad El-Solh, Beirut 1107 2020, Lebanon.
Rationale And Objectives:
To investigate radiation dose reduction during whole body fluorodeoxyglucose (18F-FDG) positron emission tomographic (PET)/computed tomography (CT) by employing weight-based protocols.
Materials And Methods:
One thousand and twenty-eight patients were referred for 18F-FDG PET/CT study with one of two protocols: conventional protocol I; 120 kVp, 120 mAs, 0.5 second rotation time, pitch 0.8 mm/rot across all body weights; four-tier body weight protocol II all used 140 kVp, 0.75 seconds rotation time and pitch 0.8 mm/rot: Protocol A (≤60 kg): 35 mAs, Protocol B (61-80 kg): 50 mAs, Protocol C (81-100 kg): 65 mAs, and Protocol D: (>101 kg): 100 mAs. All protocols employed tube current modulation. Quantitative and qualitative image visual grading characteristics assessed image quality.
Results:
Patient demographics demonstrated no significant difference between each protocol except for patient weight in weight protocol IIB (p < 0.009). Mean effective dose in all protocols were significantly lower in Protocol B compared to A (p < 0.009). Contrast-to-noise ratio demonstrated no differences between each protocol (p < 0.21) except for weight protocol in protocol IIA (<60 kg, p = 0.035) with the visual grading characteristics demonstrating preference over protocol II compared to I.
Conclusion:
Significant reduction in radiation dose can be achieved using patient-specific body weight-based protocols during whole-body 18F-FDG PET/CT without compromising image quality when employing weight-based protocols.
Related Concept Videos
Reduced Mass Coordinates: Isolated Two-body Problem
Biological Effects of Radiation
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Body Temperature
Body Temperature
The average body temperature is approximately 37°C (98.6°F) and typically ranges from 36.1–37.2°C...
Centroid of a Body
For a homogeneous body with constant density, the centroid can usually be found using equations representing a balance of the moments of the body's volume. If the...