Related Experiment Video
Updated: Feb 6, 2026

Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
Automated Segmentation of Bioresorbable Vascular Scaffold Struts in Intracoronary Optical Coherence Tomography Images
Junedh M Amrute1, Lambros Athanasiou2, Farhad Rikhtegar2
1Division of Biology and Biological Engineering California Institute of Technology Pasadena, CA, USA.
Abstract:
Bioresorbable vascular scaffolds (BVS), the next step in the continuum of minimally invasive vascular interventions present new opportunities for patients and clinicians but challenges as well. As they are comprised of polymeric materials standard imaging is challenging. This is especially problematic as modalities like optical coherence tomography (OCT) become more prevalent in cardiology. OCT, a light-based intracoronary imaging technique, provides cross-sectional images of plaque and luminal morphology. Until recently segmentation of OCT images for BVS struts was performed manually by experts. However, this process is time consuming and not tractable for large amounts of patient data. Several automated methods exist to segment metallic stents, which do not apply to the newer BVS. Given this current limitation coupled with the emerging popularity of the BVS technology, it is crucial to develop an automated methodology to segment BVS struts in OCT images. The objective of this paper is to develop a novel BVS strut detection method in intracoronary OCT images. First, we preprocess the image to remove imaging artifacts. Then, we use a K-means clustering algorithm to automatically segment the image. Finally, we isolate the stent struts from the rest of the image. The accuracy of the proposed method was evaluated using expert estimations on 658 annotated images acquired from 7 patients at the time of coronary arterial interventions. Our proposed methodology has a positive predictive value of 0.93, a Pearson Correlation coefficient of 0.94, and a F1 score of 0.92. The proposed methodology allows for rapid, accurate, and fully automated segmentation of BVS struts in OCT images.
More Related Videos
11:12Nondestructive Monitoring of Degradable Scaffold-Based Tissue-Engineered Blood Vessel Development Using Optical Coherence Tomography
Published on: October 3, 2018
09:56Automated 3D Optical Coherence Tomography to Elucidate Biofilm Morphogenesis Over Large Spatial Scales
Published on: August 21, 2019
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Seedless Vascular Plants
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...