Iterative scatter correction for grid-less skeletal radiography allows improved image quality equal to an antiscatter
Christoph G Lisson1, Catharina S Lisson1, Sebastian Kleiner1
11 Department of Diagnostic and Interventional Radiology, University Hospital of Ulm, Ulm, Germany.
Background:
Iterative scatter correction (ISC) is a new technique applicable to plain radiography; comparable to iterative reconstruction for computed tomography, it promises dose reduction and image quality improvement. ISC for bedside chest X-rays has been applied and evaluated for some time and has recently been commercially offered for plain skeletal radiography.
Purpose:
To analyze the potential of ISC for plain skeletal radiography with regard to image quality improvement, dose reduction, and replacement for an antiscatter grid.
Material And Methods:
A total of 385 radiographs with different imaging protocols of the pelvis and cervical spine were acquired from 20 body donors. Radiographs were rated by four radiologists. Ratings were analyzed with visual grading characteristics (VGC) analysis. The area under the VGC curve was used as a measure of difference in image quality.
Results:
Without ISC, the grid-less images were rated significantly worse than their grid-based counterparts (0.389, P = 0.005); adding ISC made image quality equal (0.498; P = 0.963). In grid-less imaging, reduction of dose by 50% led to significant image quality impairment (0.415, P = 0.001); this was fully counterbalanced when ISC was added (0.512; P = 0.588).
Conclusion:
ISC for plain skeletal radiography has the ability to replace the antiscatter grid without image quality impairment, to improve image quality in grid-less imaging, and to reduce patient radiation dose by 50% without substantial loss in image quality.
Related Concept Videos
One-Way ANOVA: Equal Sample Sizes
Different sample means can result in different values for the variance estimate: variance between samples. This is because the variance between samples is calculated as the product of the sample size and the variance between the...
Bioavailability Study Design: Single Versus Multiple Dose Studies
Electric Field of Two Equal and Opposite Charges
A separation of the positive and negative charges can lead to a weak, remnant effect of the positive and negative charges. The expectation is that the more the distance between the positive and...
Distance Corrections
Power Factor Correction
Heart Failure VI: Adjunct Therapies