Quadruple Semitendinosus Graft Construct With Double Cortical Suspensory Fixation for Anterior Cruciate Ligament
Carla Alexandra Madaíl1,2,3, Maria de Fátima Vaz4,5, Pedro Miguel Amaral4,5
1Orthopedics and Trauma Department of Hospital de S. Francisco Xavier, Centro Hospitalar de Lisboa Ocidental Estrada Forte do Alto do Duque, 1449-005, Lisbon, Portugal. carlamadaill19@gmail.com.
Abstract:
The purpose of this study was to evaluate the biomechanical properties of a graft construct with quadrupled Semitendinosus and two cortical buttons with adjustable loops concerning elongation, stiffness and resistance. A total of 15 fresh human cadaveric semitendinosus tendons were quadrupled over the two adjustable loops and stitched at the tibial tip with a cerclage type suture. They underwent pre-tensioning at 300 N for 2 minutes followed by cyclic loading (1000 cycles between 50-250 N) and finally a load-to-failure test. Statistical analysis was performed using SPSS Statistics software and groups were compared using a paired t-test, with a significance level set at α = 0.05. Graft construct elongation after pre-tensioning at 300 N was 12.8 mm (9.3 mm-16.5 mm) and mean cyclic elongation 0.4 mm (0.2 mm-0.9 mm), considered significant (p < 0,001). The resistance and stiffness values were respectively 849.46 N (649.30 N-1027.90 N) and 221.49 N (178,30 N - 276.10 N). Quadruple ST graft construct using two cortical buttons and adjustable loops showed a high stiffness and resistance with a very low elongation after cycling.
Related Concept Videos
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Muscles of the Anterior Neck
Carbon-dioxide Fixation
Fixing Double-strand Breaks
Fixing Double-strand Breaks
Reconstruction of Signal using Interpolation