Impact of Heart Rate on Pulsatile Hemodynamic Performance in a Neonatal ECG-Synchronized ECLS System
Shigang Wang1, Morgan K Moroi1, Madison Force1
1Penn State Health Pediatric Cardiovascular Research Center, Department of Pediatrics, Penn State Milton S. Hershey Medical Center, Penn State College of Medicine, Penn State Health Children's Hospital, Hershey, PA, USA.
Abstract:
The experimental circuit consisted of an i-cor diagonal pump, a Medos Hilite 800 LT oxygenator, an 8Fr Biomedicus arterial cannula, a 10Fr Biomedicus venous cannula, and six feet of 1/4 in ID tubing for arterial and venous lines. The circuit was primed with lactated Ringer's solution and packed red blood cells (hematocrit 40%). Trials were conducted at various heart rates (90, 120, and 150 bpm) and flow rates (200, 400, and 600mL/min) under nonpulsatile and pulsatile mode with pulsatile amplitudes of 1000-4000rpm (1000 rpm increments). Real-time pressure and flow data were recorded for analysis. The i-cor pump was capable of creating nonpulsatile and electrocardiography (ECG)-synchronized pulsatile flow, and automatically reducing pulsatile frequency by increasing the assist ratio at higher heart rates. Reduced pulsatile frequency led to lower hemodynamic energy generation but did not affect circuit pressure drop. Pulsatile flow delivered more hemodynamic energy to the pseudopatient when compared with nonpulsatile flow. The pump generated more hemodynamic energy with higher pulsatile amplitudes. The i-cor pump can automatically adjust the pulsatile assist ratio to create pulsatile flow at higher heart rates, although this caused some hemodynamic energy loss. Compared with nonpulsatile flow, pulsatile flow generated and transferred more hemodynamic energy to the neonate during ECLS (200-600mL/min), especially at high pulsatile amplitudes and low flow rates.
More Related Videos
08:12Calculating Heart Rate Variability from ECG Data from Youth with Cerebral Palsy During Active Video Game Sessions
Published on: June 5, 2019
09:11Implantation of Radiotelemetry Transmitters Yielding Data on ECG, Heart Rate, Core Body Temperature and Activity in Free-moving Laboratory Mice
Published on: November 21, 2011
Related Concept Videos
Regulation of Heart Rates
The SNS increases heart rate through the release of norepinephrine and epinephrine, which act on beta-1 adrenergic receptors in the heart. This action increases the rate of depolarization in the sinoatrial (SA) node, the heart's...
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Impact of Groups on Groups
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...