Significant Differences in Debris Captured by the Sentinel Dual-Filter Cerebral Embolic Protection During
Julia Seeger1, Renu Virmani2, Maria Romero2
1Department of Internal Medicine II, Cardiology, University of Ulm, Ulm, Germany.
Objectives:
The aim of this study was to evaluate the debris captured by the Claret Sentinel cerebral embolic dual-filter protection device during transfemoral transcatheter aortic valve replacement (TAVR) with different valve types.
Background:
Risk for embolization of debris during TAVR may vary by TAVR device.
Methods:
The filters of 100 consecutive patients were collected and captured debris was analyzed by histopathology and histomorphometry. Three valve types were implanted: the balloon-expandable Edwards SAPIEN 3 (n = 42), the self-expandable Medtronic Evolut R (n = 35), and the mechanically implantable Boston Scientific Lotus (n = 23).
Results:
Among the 3 groups there was no difference in baseline data, including Society of Thoracic Surgeons score for mortality, calcification, or pre-dilation. The type of captured debris did not differ among the 3 valve types in the proximal or distal filter. With the balloon-expandable valve, there were significantly more patients with large debris measuring ≥1,000 μm. The number of particles in the proximal filter was significantly lower with the Lotus (89.8 ± 106.3) compared with the Evolut R (187.3 ± 176.9) and Edwards SAPIEN 3 (172.3 ± 133.5) valves (p = 0.035). Total tissue area in the proximal filter was significantly smaller for the Lotus compared with the other 2 valve types (7.1 ± 6.3, 20.1 ± 19.0, and 21.3 ± 15.1 mm2; p = 0.0014). In contrast, for the distal filter, there were no differences with respect to valve type for total tissue area, particle size, and number of particles.
Conclusions:
A significant difference was observed in the size and number of captured tissue particles with the double-filter embolic protection device among different valve types in patients undergoing TAVR. The largest particles were observed in patients treated with a balloon-expandable valve.
More Related Videos
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Pulmonary Embolism I: Introduction