X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Radiation dose reduction for musculoskeletal computed tomography of the pelvis with preserved image quality
Elaina Zabak1, Hythem Omar1, Ethan Boothe2
1UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX, 75390-9178, USA.
Objective:
To analyze the impact of pelvic computed tomography (CT) technique optimization on estimated dose and subjective and objective image quality.
Materials And Methods:
An institutional review board (IRB)-approved retrospective records review was performed with waived informed consent. Five CT scanners (various manufacturers/models) were standardized to match the lowest dose profile on campus via subjective assessment of clinical images by experienced musculoskeletal radiologists. The lowest dose profile had previously been established through image assessment by experienced musculoskeletal radiologists after a department-wide radiation dose reduction initiative. A consecutive series of 60 pre- and 59 post-optimization bony pelvis CTs were analyzed by two residents, who obtained signal-to-noise ratio for femoral cortex and marrow, gluteus medius muscle, and subcutaneous and visceral fat in a standardized fashion. Two blinded attending radiologists ranked image quality from poor to excellent.
Results:
Pre- and post-optimization subjects exhibited no difference in gender, age, or BMI (p > 0.2). Mean CT dose index (CTDIvol) and dose-length product (DLP) decreased by approximately 45%, from 39± 14 to 18± 12 mGy (p < 0.0001) and 1,227± 469 to 546± 384 mGy-cm (p < 0.0001). Lower body mass index (BMI) was associated with a larger dose reduction and higher BMI with higher DLP regardless of pre- or post-optimization examination. Inter-observer agreement was 0.64-0.92 for SNR measurements. Cortex SNR increased significantly for both observers (p < 0.02). Although qualitative image quality significantly decreased for one observer (p < 0.01), adequate mean quality (3.3 out of 5) was maintained for both observers.
Conclusion:
Subjective and objective image quality for pelvic CT examination remains adequate, despite a substantially reduced radiation dose.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Biological Effects of Radiation
Veins of the Abdomen and Pelvis
The inferior vena cava is fed by numerous smaller veins. The lumbar veins, for instance, drain the posterior abdominal wall, emptying both directly into the inferior vena cava and into the...
Oxidation-Reduction Reactions
Radiation: Applications
The average...