Commutability of external quality assessment materials for serum sodium and potassium measurements
Ying Yan1,2, Bingqing Han1,2, Haijian Zhao1,2
1Beijing Hospital, National Center of Gerontology, National Center for Clinical Laboratories and Beijing Engineering Research Center of Laboratory Medicine, Dongdan, Beijing, P.R. China.
Abstract:
Background The commutability of electrolyte trueness verification materials (ETVs) and commercial general chemistry materials (GCs) was evaluated to investigate their suitability for use in an external quality assessment (EQA) program for serum sodium and potassium measurements. Methods Eighty fresh individual human samples (40 for sodium measurements and 40 for potassium measurements), six ETVs and three GCs were analyzed by five routine methods (validated methods) and by inductively coupled plasma mass spectrometry reference methods (comparative methods) for the determination of sodium and potassium. The commutability was analyzed according to Clinical and Laboratory Standards Institute (CLSI) EP14-A3 protocol and difference in bias approach, respectively. The linearity, bias and imprecision of the routine methods were also assessed according to CLSI guidelines. Results According to EP14-A3 protocol, ETVs were commutable for all assays, and GCs were commutable for 3/5 assays for sodium. ETVs were commutable in most assays except Cobas C501, while GCs showed no commutability except in case of AU5821 for potassium. According to a difference in bias approach, the commutability of ETVs was inconclusive for most routine assays for both sodium and potassium, and GCs were inconclusive for sodium and non-commutable for potassium in most routine assays. The routine methods exhibited excellent linearities and precisions. The majority and minority of relative biases between the routine and reference methods were beyond the bias limits for sodium and potassium, respectively. Conclusions Superiority in the commutability of ETVs over GCs was observed among the sodium and potassium assays whichever evaluation approach was applied.
Related Concept Videos
Regulation of Sodium and Potassium
Sodium Regulation
Sodium ions make up approximately 90% of extracellular cations, with a normal blood plasma concentration of 136–148 mEq/L. A decrease in blood volume and pressure triggers the release of renin from granular cells in the juxtaglomerular complex (JGC), primarily...
Roles of Electrolytes: Sodium and Potassium
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
Internal and External Forces
External and Internal Respiration
External Anatomy of the Kidney
The kidneys are located in the retroperitoneal space on either side of the vertebral column, protected posteriorly by the 11th and 12th ribs. The right kidney sits slightly lower than the left owing to the presence of the liver...