Bubble-free Gonioscopy
Frederick M Kapetansky1, Nathaniel B Dusseau
1Department of Ophthalmology and Visual Science, The Ohio State University Wexner Medical Center, Havener Eye Institute, Columbus, OH.
A new 2-mirror goniolens successfully eliminated the need for coupling gel in 84% of cases. This innovation aims to improve glaucoma diagnosis and treatment by preventing air bubbles during gonioscopy.
Area of Science:
- Ophthalmology
- Glaucoma Diagnosis
- Ophthalmic Device Development
Background:
- Accurate gonioscopy is crucial for glaucoma diagnosis and treatment.
- The Goldmann 3-mirror lens, the current standard, requires coupling gel and can trap air bubbles.
- An ideal gonioscopy lens would eliminate the need for gel and prevent air bubble formation.
Purpose of the Study:
- To evaluate a prototype 2-mirror goniolens designed to eliminate the need for coupling gel and prevent air bubble trapping.
- To assess the efficacy of the new lens in clinical practice.
Main Methods:
- A prototype 2-mirror goniolens with specific curvature and diameter was used.
- Bilateral keratometer measurements were taken for 69 patients.
- The lens was applied to anesthetized eyes, and outcomes (air bubble trapping) were recorded for 64 patients.
Main Results:
- A total of 64 patients were evaluated with the prototype lens.
- 84.38% (54/64) of eyes were successfully evaluated without air bubble trapping.
Conclusions:
- The prototype 2-mirror goniolens successfully eliminated the need for coupling gel and prevented air bubbles in 84.38% of cases.
- These findings support the development of novel bubble-free goniolenses for procedures like laser trabeculoplasty.
Related Concept Videos
Excess Pressure Inside a Drop and a Bubble
Phase Transitions: Vaporization and Condensation
Physical Properties Affecting Solubility
As for any solution, the solubility of a gas in a liquid is affected by the attractive intermolecular forces between solute and solvent species. Unlike solid and liquid solutes, however, there is no solute-solute intermolecular attraction to overcome when a gaseous solute dissolves in a liquid solvent since the atoms or molecules comprising a gas are far separated and experience negligible interactions. Consequently, solute-solvent interactions are the sole...
Replication in Eukaryotes
Bacterial RNA Polymerase
In most genes, the transcription site is a single base present upstream of the coding sequence. Though RNAP is a catalytically efficient enzyme, it does not recognize...
Qualitative Analysis
For instance, group IV...