Defibrillator Design and Usability May Be Impeding Timely Defibrillation
Background:
Timely defibrillation is the only rhythm-specific therapy proven to increase survival to hospital discharge following cardiac arrest secondary to ventricular tachyarrhythmia. Delayed defibrillation occurs in more than 30% of this population. A study was conducted to test the hypothesis that unintuitive defibrillator design and lack of usability are barriers to timely defibrillation, as measured by time to defibrillation and the proportion of defibrillations delivered within 2 minutes.
Methods:
A mixed-methods (qualitative and quantitative) prospective usability study was performed to evaluate the use of a defibrillator in a simulated hospital environment. Participants were asked to perform two simulated tasks typical of in-hospital cardiac arrest care: defibrillation and synchronized cardioversion.
Results:
The average time to defibrillation was 4 minutes 21 seconds. Only 9.1% of participants (2/22) performed a defibrillation within 2 minutes. Participants had difficulty with several aspects of defibrillator use, including attaching the hands-free defibrillator electrode pads and selecting an appropriate display. Participants rated defibrillator design 4.2 ± 1.8 (mean, standard deviation) on a perceived usability scale (1 = "poorly designed"; 9 = "perfectly designed").
Conclusion:
Most participants were unable to perform a simulated defibrillation within 2 minutes. This delay in defibrillation was likely at least partially the result of poor defibrillator design and lack of usability. Expert observation and participant feedback were largely congruent in terms of which aspects of defibrillator design do not suit the end user. Modification of future defibrillator design, along with improved education and training, may result in more timely defibrillation.
Related Concept Videos
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Group Design
Factorial Design
Series Impedances: Three-Phase Line
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...
Bus Impedance Matrix
In the first circuit, all machine voltage sources are short-circuited, leaving only the prefault voltage source at the fault location. The positive-sequence bus impedance matrix can be determined by solving the nodal equations,...