Second-generation bi-cruciate stabilized total knee system has a lower reoperation and revision rate than its
Bernhard Christen1, Branko Kopjar2
1CHRISTENORTHO AG, Schaenzlistrasse 39, 3000, Bern 25, Switzerland. b.christen@christenortho.ch.
Introduction:
Total knee arthroplasty (TKA) can provide pain relief and good long-term results. However, nearly 30% of post-surgical patients are unsatisfied due to persistent pain and functional deficits. A second-generation bi-cruciate stabilized TKA device has a post-cam mechanism with an asymmetric femoral component, a polyethylene insert, and a medially concave and laterally convex shape. The device is designed to provide guided motion, and thus improve knee kinematics by more closely approximating a normal knee. The aim of this study was to evaluate early complication and revision rates of the second-generation device and to compare its clinical performance to the first-generation device.
Materials And Methods:
In this retrospective, longitudinal, non-concurrent cohort study, 140 TKAs were performed using the second-generation device on 131 patients from 2012 to 2016, and 155 TKAs were performed using the first-generation device on 138 patients from 2009 to 2012. Primary outcomes were occurrence of revisions and reoperations.
Results:
There were 31 reoperations [3.21 per 100 observed component years (OCY)] in 22 (2.28 per 100 OCY) TKAs in the first-generation device cohort compared to five reoperations (1.92 per 100 OCY) in four TKAs (1.54 per 100 OCY) in the second-generation device cohort. The adjusted hazard ratio (HR) was 3.50 (P = 0.0254). There were 21 revisions (2.17 per 100 OCY) in 16 (1.66 per 100 OCY) TKAs in the first-generation device cohort, compared to only three revisions (1.15 per 100 OCY) in two TKAs (0.77 per 100 OCY) in the second-generation device cohort. The adjusted HR was 4.16 (P = 0.0693).
Conclusion:
The improved design of the second-generation device appears to be associated with a lower risk of reoperation and revision compared to that of the first-generation device.
Level Of Evidence:
III.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Nuclear Stability
To hold positively charged protons together...
Speciation Rates
RNA Stability
Stability
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...