Contrast-to-Noise Ratio Optimization in Coronary Computed Tomography Angiography: Validation in a Swine Model
Logan Hubbard1, Shant Malkasian1, Yixiao Zhao1
1Department of Radiological Sciences, Medical Sciences I, B-140, University of California, Irvine, Irvine, CA 92697.
Rationale And Objectives:
The accuracy of coronary computed tomography (CT) angiography depends upon the degree of coronary enhancement as compared to the background noise. Unfortunately, coronary contrast-to-noise ratio (CNR) optimization is difficult on a patient-specific basis. Hence, the objective of this study was to validate a new combined diluted test bolus and CT angiography protocol for improved coronary enhancement and CNR.
Materials And Methods:
The combined diluted test bolus and CT angiography protocol was validated in six swine (28.9 ± 2.7 kg). Specifically, the aortic and coronary enhancement and CNR of a standard CT angiography protocol, and a new combined diluted test bolus and CT angiography protocol were compared to a reference retrospective CT angiography protocol. Comparisons for all data were made using box plots, t tests, regression, Bland-Altman, root-mean-square error and deviation, as well as Lin's concordance correlation.
Results:
The combined diluted test bolus and CT angiography protocol was found to improve aortic and coronary enhancement by 26% and 13%, respectively, as compared to the standard CT angiography protocol. More importantly, the combined protocol was found to improve aortic and coronary CNR by 29% and 20%, respectively, as compared to the standard protocol.
Conclusion:
A new combined diluted test bolus and CT angiography protocol was shown to improve coronary enhancement and CNR as compared to an existing standard CT angiography protocol.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Reliability and Validity
Poisson's Ratio
Odds Ratio
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...