High-resolution mapping of pulmonary vein potentials improved the successful pulmonary vein isolation using small
Chin-Yu Lin1, Abigail Louise D Te2, Yenn-Jiang Lin3
1Heart Rhythm Center, Division of Cardiology, Department of Medicine, Taipei Veterans General Hospital, Taiwan; Institute of Clinical Medicine, National Yang-Ming University School of Medicine, Taipei, Taiwan; Department of Internal Medicine, Taipei Veterans General Hospital, Yuan-Shan Branch, I-Lan, Taiwan.
Background:
Intracardiac electrogram recording is influenced by the electrode size and inter-electrode spacing. Smaller electrodes with a closer inter-electrode spacing may improve the mapping resolution and outcome.
Methods:
Substrate mapping of the left atrium and residual pulmonary vein (PV) potentials during sinus rhythm was sequentially performed using a 3.5-mm electrode tip catheter and a 1-mm electrode multielectrode catheter in 33 patients (Group 1) that underwent repeat atrial fibrillation (AF) procedures. PV gap identification and electrophysiological characteristics were compared. Arrhythmia freedom was compared with a propensity matched (1:2) control group (66 patients, Group 2) undergoing repeat AF procedures guided by wide inter-electrode spacing catheter.
Results:
In the Group 1 patients, the total area of residual PV potentials measured using the 1-mm catheter was larger than that measured by the 3.5-mm catheter. Overall 1.97 ± 0.59 (1-3) and 1.49 ± 0.62 (1-3) PVs were identified by the 1-mm electrode and 3.5 mm catheters, respectively (P = 0.02). The gaps not identified by the 3.5 mm catheter had a smaller width and lower voltage. Radiofrequency catheter ablation in the areas with residual PV potentials identified by the 1-mm catheter resulted in complete electrical isolation of the PVs. Arrhythmia freedom at one year of follow-up was achieved in 26 of 33 (78.8%) patients in Group 1, which was significantly higher than the matched control group (33/66 [50%], P < 0.05).
Conclusion:
In the patients with a previous PV isolation, mapping with small, closely spaced electrodes can increase the detection rate of residual PV potentials and improve the outcome.
More Related Videos
Related Concept Videos
Standard Electrode Potentials
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...
Veins of Thorax
The azygos vein, positioned just right of the midline and anterior to the vertebral column, begins at the junction of the right ascending lumbar and subcostal veins, terminating in the superior vena cava. This vein drains blood from the right side of the thoracic wall, thoracic viscera, and posterior abdominal wall.
The...
Veins of Lower Limbs
Formed by the union of the medial and lateral plantar veins, the posterior tibial vein, rising through the calf muscle, assimilates the fibular vein. The anterior tibial vein, a superior extension of the foot's dorsalis pedis vein, merges with the posterior tibial vein at the...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Pulmonary Tuberculosis I
Causative Organism
The primary infectious agent causing tuberculosis is Mycobacterium tuberculosis, a slow-growing, acid-fast, aerobic rod that exhibits sensitivity to heat and ultraviolet light. Instances of Mycobacterium bovis and Mycobacterium avium contributing to the development of TB infection are rare.
Mode of...