Comparison of incision size and intraocular lens performance after implantation with three preloaded systems and one manual delivery system
Related Experiment Video
Updated: Feb 5, 2026

Simulating the Mechanics of Lens Accommodation via a Manual Lens Stretcher
Published on: February 23, 2018
Javier Mendicute1, Thierry Amzallag2, Lixin Wang3
1Department of Ophthalmology, Hospital Universitario Donostia, San Sebastián, Spain, mendicutejavier@gmail.com.
Purpose:
To compare corneal incision size and intraocular lens (IOL) performance/behavior following implantation with the following delivery systems: system U (UltraSert®), system S (Hoya iSert® 250/251), system T (Tecnis® iTec), and a manual system (Monarch® III Delivery System).
Setting:
Six study sites (four in Spain and two in France).
Design:
Prospective, multicenter, parallel-group, randomized, subject-masked, postmarket clinical study.
Materials And Methods:
Subjects were enrolled based on predetermined inclusion/exclusion criteria. The effectiveness end points compared corneal incision size and enlargement after IOL implantation (day of surgery) among all delivery systems. Exploratory end points included mean enlargement of corneal incision size, rates of trapped trailing haptic, IOL adherence to the plunger tip, nozzle tip splitting, and mean surgically induced astigmatism (SIA) at postoperative visit.
Results:
One hundred and nine subjects participated in the study. The mean corneal incision size following IOL implantation was 2.35±0.019 mm for system U, 2.47±0.016 mm for system T, 2.54±0.019 mm for system S, and 2.49±0.011 mm for the manual system. There were five instances of trapped trailing haptic (all system T group, N=26), one instance of IOL adherence to the plunger tip (system S group, N=26), and six instances of nozzle tip splitting (all system S group, N=26). System U had the least SIA (postoperative Day 1) (SIA Centroid = 0.10 diopters [axis: 83.06°]).
Conclusion:
Preloaded delivery system U supported the completion of surgery with the smallest incision size, the least SIA (postoperative Day 1), and no trapped trailing haptics or nozzle tip splitting compared to two other preloaded systems and one manual system.
More Related Videos
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Cell Size
Surface Area
Cells can take in nutrients and water via diffusion through the plasma membrane itself or through specific channels in the membrane. The area of the membrane surrounding...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Second Order systems II
Evolutionary Relationships through Genome Comparisons
First Order Systems
When a first-order system is subjected to a unit-step input, its response is characterized by its transfer function. By applying the Laplace transform of the unit-step input to the transfer function, expanding the...