A High-performance Compact Photoacoustic Tomography System for In Vivo Small-animal Brain Imaging
Published on: June 21, 2017
Adjustable photoacoustic tomography probe improves light delivery and image quality
Gurneet S Sangha1, Nicholas J Hale1, Craig J Goergen1,2
1Weldon School of Biomedical Engineering, Purdue University, 206 S. Martin Jischke Dr., West Lafayette, IN 47907, USA.
Abstract:
One cause for suboptimal photoacoustic tomography (PAT) penetration depth is attenuation of incident light by soft tissue. To better understand this problem, we investigated the effects of illumination fiber optic bundle geometry on PAT penetration depth and signal-to-noise ratio. An adjustable, motorized PAT probe was used to reduce probe-skin reflection artifacts and improve light distribution in the image acquisition plane by tuning fiber orientation. We validated our motorized PAT probe through Monte Carlo simulations and ex vivo imaging of a tissue mimicking phantom, and in vivo imaging of murine periaortic fat. Overall, our ex vivo results showed a several millimeter improvement in penetration depth and in vivo results showed a >62% increase in lipid signal-to-noise ratio. Our PAT probe also utilized a 7-μm aluminum filter to block in vivo probe-skin reflection artifacts. Together, these findings showed the importance of optimizing illumination geometry to enhance PAT image quality.
More Related Videos
11:21Integrated Photoacoustic Ophthalmoscopy and Spectral-domain Optical Coherence Tomography
Published on: January 15, 2013
06:45Author Spotlight: Integrated Photoacoustic, Ultrasound, and Angiographic Tomography (PAUSAT) for NonInvasive Whole-Brain Imaging of Ischemic Stroke
Published on: June 2, 2023
Related Concept Videos
The Anchoring-and-Adjustment Heuristic
Adjusting a Traverse
Imaging Studies III: Computed Tomography
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Photoreceptors and Plant Responses to Light