Artificial ventilation during transport: A randomized crossover study of manual resuscitators with comparison to
Malcolm J Lucy1, Jonathan J Gamble1, Andrew Peeling1
1Department of Anesthesiology, University of Saskatchewan, Saskatoon, SK, Canada.
Background:
Positive-pressure ventilation in critically ill patients is commonly administered via a manual resuscitation device or a mechanical ventilator during transport. Our group previously compared delivered ventilation parameters between a self-inflating resuscitator and a flow-inflating resuscitator during simulated in-hospital pediatric transport. However, unequal group access to inline pressure manometry may have biased our results. In this study, we examined the performance of the self-inflating resuscitator and the flow-inflating resuscitator, both equipped with inline manometry, and several mechanical ventilators to deliver prescribed ventilation parameters during simulated pediatric transport.
Methods:
Thirty anesthesia providers were randomized to initial resuscitator device used to hand ventilate a test lung. The resuscitators studied were a Jackson-Rees circuit (flow-inflating resuscitator) or a Laerdal pediatric silicone resuscitator (self-inflating resuscitator), both employing manometers. The scenario was repeated using several mechanical transport ventilators (Hamilton-T1, LTV® 1000, and LTV® 1200). The primary outcome was the proportion of total breaths delivered within the predefined target PIP/PEEP range (30 ± 3, 10 ± 3 cm H2 O).
Results:
The Hamilton-T1 outperformed the other ventilators for breaths in the recommended range (χ2 = 2284, df = 2, P < .001) and with no breaths in the unacceptable range (χ2 = 2333, df = 2, P < .001). Hamilton-T1 also outperformed all human providers in proportion of delivered acceptable and unacceptable breaths (χ2 = 4540, df = 3, P < .001 and χ2 = 639, df = 3, P < .001, respectively). Compared with the flow-inflating resuscitator, the self-inflating resuscitator was associated with greater odds of breaths falling outside the recommended range (Odds ratio (95% CI): 1.81 (1.51-2.17)) or unacceptable (Odds ratio (95% CI): 1.63 (1.48-1.81)).
Conclusion:
This study demonstrates that a majority of breaths delivered by manual resuscitation device fall outside of target range regardless of provider experience or device type. The mechanical ventilator (Hamilton-T1) outperforms the other positive-pressure ventilation methods with respect to delivery of important ventilation parameters. In contrast, 100% of breaths delivered by the LTV 1200 were deemed unacceptable.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: