Analysis of differences in intraocular pressure evaluation performed with contact and non-contact devices
Michele Lanza1, Michele Rinaldi2, Ugo Antonello Gironi Carnevale2
1Multidisciplinary Department of Medical, Surgical and Dental Sciences, Università della Campania, Luigi Vanvitelli, Via de Crecchio 16, 80100, Naples, Italy. mic.lanza@gmail.com.
Background:
To evaluate differences of intraocular pressure (IOP) measurements performed with Goldmann applanation tonometer (GAT), dynamic contour tonometer (DCT), rebound tonometry (RT), Ocular Response Analyzer (ORA) and Corvis ST (CST) in eyes screened for refractive surgery.
Methods:
One eye, only the right one, of 146 patients was included in this study. Each participant was submitted to a corneal analysis with Scheimpflug camera and IOP evaluation with GAT, DCT, RT, ORA and CST. Differences in IOP values obtained thanks to each instruments were compared and then correlations between these discrepancies and morphological features such as mean keratometry (MK) and central corneal thickness (CCT) provided by Pentacam were studied. Software used to run statistical evaluations was SPSS, version 18.0.
Results:
Study participants had a mean age of 33.1 ± 9.2 years old. IOP values observed in this study were 15.97 ± 2.47 mmHg (GAT), 17.55 ± 2.42 mmHg (DCT), 17.49 ± 2.08 mmHg (RT), 18.51 ± 2.59 mmHg (ORA) and 18.33 ± 2.31 mmHg (CST). The mean CCT was 560.23 ± 31.00 μm, and the mean MK was 43.33 ± 1.35 D. GAT provided significant lower values in comparison to all other devices. DCT and RT gave significantly lower intermediate IOP values than those measured with ORA and CST. All the IOP measures and the differences between devices were significantly correlated with CCT.
Conclusions:
According to our data, although our findings should be confirmed in further studies, GAT tonometer cannot be used interchangeably with DCT, RT, ORA and CST.
More Related Videos
Related Concept Videos
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Transmission-based Precautions I: Contact, Enteric, and Droplets
Contact Precautions:
Contact precautions are the measures taken to prevent the transmission of infectious agents, especially epidemiologically important microorganisms such as MRSA or influenza, primarily transmitted through direct or indirect contact with an...
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Performing a Simple Data Analysis using MS-Excel Function
SUM: This function calculates the total sum of a range of values. It's the foundation for aggregating data, essential for determining overall trends and totals in datasets.
AVERAGE: It computes the mean value of a given set of numbers, providing a quick insight into the central...
Electric Potential and Potential Difference
When a test charge moves from the initial to the final position, the electric potential difference between those positions is defined as the ratio of the change in the potential energy to the charge on the...