Multienergy element-resolved cone beam CT (MEER-CBCT) realized on a conventional CBCT platform
Chenyang Shen1, Bin Li1,2, Yifei Lou3
1Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX, 75390, USA.
Purpose:
Cone beam CT (CBCT) has been widely used in radiation therapy. However, its main application is still to acquire anatomical information for patient positioning. This study proposes a multienergy element-resolved (MEER) CBCT framework that employs energy-resolved data acquisition on a conventional CBCT platform and then simultaneously reconstructs images of x-ray attenuation coefficients, electron density relative to water (rED), and elemental composition (EC) to support advanced applications.
Methods:
The MEER-CBCT framework is realized on a Varian TrueBeam CBCT platform using a kVp-switching scanning scheme. A simultaneous image reconstruction and elemental decomposition model is formulated as an optimization problem. The objective function uses a least square term to enforce fidelity between x-ray attenuation coefficients and projection measurements. Spatial regularization is introduced via sparsity under a tight wavelet-frame transform. Consistency is imposed among rED, EC, and attenuation coefficients and inherently serves as a regularization term along the energy direction. The EC is further constrained by a sparse combination of ECs in a dictionary containing tissues commonly existing in humans. The optimization problem is solved by a novel alternating-direction minimization scheme. The MEER-CBCT framework was tested in a simulation study using an NCAT phantom and an experimental study using a Gammex phantom.
Results:
MEER-CBCT framework was successfully realized on a clinical Varian TrueBeam onboard CBCT platform with three energy channels of 80, 100, and 120 kVp. In the simulation study, the attenuation coefficient image achieved a structural similarity index of 0.98, compared to 0.61 for the image reconstructed by the conventional conjugate gradient least square (CGLS) algorithm, primarily because of reduction in artifacts. In the experimental study, the attenuation image obtained a contrast-to-noise ratio ≥60, much higher than that of CGLS results (~16) because of noise reduction. The median errors in rED and EC were 0.5% and 1.4% in the simulation study and 1.4% and 2.3% in the experimental study.
Conclusion:
We proposed a novel MEER-CBCT framework realized on a clinical CBCT platform. Simulation and experimental studies demonstrated its capability to simultaneously reconstruct x-ray attenuation coefficient, rED, and EC images accurately.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Shear on the Horizontal Face of a Beam Element
Elements and Compounds
Elements
Elements are classified as atomic or molecular based on the nature of their basic units. They are unique forms of matter with specific chemical and physical properties that cannot break down into smaller substances by ordinary chemical reactions. There...
Periodic Classification of the Elements
Classification of Elements and Compounds
Compounds are pure substances composed of two or more elements in fixed, definite proportions. Compounds are classified as ionic or molecular (covalent) based on the bonds...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...