Delivery of Antibodies into the Murine Brain via Convection-enhanced Delivery
Published on: July 18, 2019
Maximising coverage of brain structures using controlled reflux, convection-enhanced delivery and the recessed step
Owen Lewis1, Max Woolley2, David E Johnson2
1School of Engineering, Queen's Building, Cardiff University, The Parade, Cardiff CF24 3AA, UK; Functional Neurosurgery Research Group, University of Bristol, School of Clinical Sciences, Learning & Research Building, Southmead Hospital, UK; Neurological Applications Department, Renishaw PLC, New Mills, Wotton-Under-Edge, Gloucestershire, GL12, UK.
Background:
The design and use of convection-enhanced delivery catheters remains an active field as clinical trials have highlighted suboptimal distribution as a contributory factor to the failure of those studies. Recent studies indicate limitations and challenges in achieving target coverage using conventional point source delivery.
New Method:
The recessed step catheter(RSC), developed by this group, does not function as a point source delivery device, but instead uses 'controlled reflux' of the infusate to a flow inhibiting recess feature. Here we investigate a range of clinically useful step lengths in agarose gel and investigate proof-of-principle in vivo(n = 5). Infusion morphology was characterised in terms of length, width and distribution volume over a range of flow rates.
Results:
For a fixed infusion volume, increases in catheter step length strongly correlated with increases in the length and volume of distribution (r>0.90, p < 0.001) whilst there were small reductions in the width of distribution (r<-0.62, p < 0.001). Step lengths below 6 mm produced spherical distributions while steps above 12 mm produced elongated distributions. Increasing peak flow rates resulted in significant reductions in distribution volume at each step length, and an increased risk of reflux beyond the step. Modifications to the infusion morphology using changes in step length were confirmed in vivo.
Conclusions:
The combination of the recessed step and the ability to adjust the step length with this catheter design make it highly suitable for tailoring the distribution volume of the infusate to meet specific morphological target volumes in the brain.
More Related Videos
Related Concept Videos
Gastroesophageal Reflux Disease I: Meaning and Pathophysiology
Conduction, Convection and Radiation: Problem Solving
In order to solve a problem related to heat transfer, first of all, the situation needs to be examined to determine the type of heat transfer involved. This could...
Gastroesophageal Reflux Disease II: Clinical Features and Management
Clinical Manifestations
GERD presents itself in a multitude of ways, with symptoms varying from person to person. The hallmark symptoms are...
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...
Self-Evaluation: Self-Enhancement and Self-Verification
Energy-releasing Steps of Glycolysis
The first energy-releasing step—the 6th step of glycolysis...