Purpose:
To evaluate the pullout strength of the all-suture anchor (ASA), based on the angles of anchor insertion and traction.
Methods:
Synthetic saw bones of 2 densities (0.16 and 0.32 g/cm3) with 3 mm thick cortical bone models were used. ASAs were inserted at 45°, 60°, 75°, or 90° and pulled at 2 angles from the surface: 45° (simulating the physiological pull of the supraspinatus) and 90° (simulating pulling out during knot tying). Five consecutive pullout tests for each insertion and traction angle combination per saw bone were conducted to evaluate the ultimate load to failure and mode of failure (80 tests total). Thereafter, 9 matched pairs of human cadaveric humeri with 2 ASA types were used (insertion angles, 45°, 75°, 90°; traction angle, 90°). Nine consecutive tests were conducted for each insertion angle and anchor type (54 tests total).
Results:
The pullout strength was significantly higher for high density- than for low-density saw bones (all P < .05). The pullout strength was higher at the 45°than at the 90° traction angle (all P < .05) and was significantly higher at the 90° and 75° than at the 45° insertion angle in both high-density saw bones and cadaveric humeri (all P < .05). However, the pullout strength was not significantly different by ASA type (all P > .05).
Conclusions:
ASA showed stronger pullout strength in higher density bones. Furthermore, it presented stronger pullout strength in the physiological traction direction of supraspinatus rather than in the knot-tying direction, consistent with the deadman theory. However, stronger pullout strength was observed in the vertically directed insertion angle, not 45°. Therefore, implanting the ASA vertically may be clinically more beneficial not only when performing knot tying during surgery, but also when the supraspinatus tendon loads the ASA postoperatively.
Clinical Relevance:
The study provides biomechanical evidence that the optimal insertion angle for an ASA is more vertical than the 45°.
Related Concept Videos
The Anchoring-and-Adjustment Heuristic
Anchoring Junctions
Lipids as Anchors
The carboxy-terminal of most of the prenylated proteins, such as Ras proteins, contains...
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...